Internal parameters table 2 

 
Par
ame
ter 

Parameter Name 

Parameter 

Function 

Description 

(Range)Factory 

Value 

LC 

password 

When  Lc=3,  it  can  view  and 
modify parameters  

 
AL2 

Second 

set 

over-temperatur
e alarm 

When  the  "secondary  set 
temperature  measurement  ≥
second  set  temperature  set 
value  +  AL2",  the  heating 
output is turned off. 

( 0.~100. ℃ )
10.0 

cT2  Second 

set 

compressor delay 

Second  set  compressor  delay 
protection time 

(0~60 minutes)

T--  heating  control 

cycle 

heating control cycle 

(1~60 seconds)

p--  heating 

proportional 
band 

time 

proportional 

action 

regulation 

(1~rH℃)15 

I--  heating integral 

time 

integral adjustment function (

1~1000 

seconds)200 

d--  heating calculus 

time 

differential 

action 

adjustment 

1~1000 

seconds)100 

 

 

 

 

Pb2  second 

set 

zeroadjustment 

correcting the error occurred 
by the sensor measurement  at 
low temperature. 
Pb2=  actual  value  - meter 
measurement 
 

( — 99.9~99.9
℃) 
    0 

 
PK2 

second set 

full-degree 

adjustment 

correcting the error occurred 
by the sensor measurement  at 
high temperature. 
PK2=1000* ( actual  value  -
meter  measurement ) /  meter 
measurement 

 
(—9.9~999) 
    0 

rL
— 

range 

lower 

limit 

the  lower  limit  of  the 
temperature measurement value

( — 150~0.0 ℃ )

—150 

rH 

range 

upper 

limit 

the  upper  limit  of  the 
temperature measurement value

(0.0~300℃) 

300 

Summary of Contents for WBLH-30-200H

Page 1: ...1 Cooling Heating Circulator WBLH 30 200H Operation Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 10 18...

Page 2: ...sponsibility of the person in charge Keep the operating instructions in a fixed position for easy reading Only specially trained staff can operate this instrument Correct operation requires the operat...

Page 3: ...working area during temperature control The decision on the work area should fully consider the danger and the use It must be placed in a non explosion proof area V Safety device standards Graded gra...

Page 4: ...ment is used as a device that provides low temperature cold source high temperature heat source Both the high and low temperatures are controlled by an intelligent controller When the temperature is t...

Page 5: ...nt ventilation air the space is more than 30cm around 2 When selecting the heat conduction medium it is necessary to pay attention to whether the upper and lower limits of its temperature the ignition...

Page 6: ...lowly At the same time ensure that no heat conduction medium overflows Protective equipment must be worn when filling such as anti corrosion chemical gloves goggles etc 6 The setting parameters to be...

Page 7: ...r The working capacity and viscosity of the heat conduction oil which is within the temperature control range of the equipment can meet the constant temperature process requirements of the load equipm...

Page 8: ...e indicators on the phase sequence protector turn green If the indicator light is flashing unlit it proves that there is failure of phase sequence phase interruption or phase sequence and open phase p...

Page 9: ...l needs The instrument model is HD 30 and the inlet and outlet port size is DN15 A valve three way is installed at the inlet of the heat conduction medium of the instrument and equipment to facilitate...

Page 10: ...may cause the heat conduction medium to penetrate the device electrical device which may cause the device to short circuit electrical failure and control system failure if the heat conduction medium i...

Page 11: ...the inlet of the heat conduction fluid at the upper end of the reactor using a stainless steel corrugated pipe connect the outlet of the heat conduction fluid of the high and low temperature circulat...

Page 12: ...urn on the Run button f Set the temperature to 150 and continue exhausting If there are a lot of bubbles during the heating process turn off the operation key and repeat step c V To open the temperatu...

Page 13: ...u can add appropriate amount of heat conduction oil When it is too high you can also discharge appropriate amount of heat conduction oil VIII To set the temperature Custom temperature setting enter th...

Page 14: ...nce the second temperature sensor PT100 thermal resistance 2 The range of product temperature measurement is 30 to 200 3 All electrical components are Schneider products the compressor is Danfoss and...

Page 15: ...set of temperature and measured value click this button to decrease the set value in the setting state and the set value can be decreased continuously by long pressing this button 4 Cycle allow butto...

Page 16: ...perature setting state the upper row of digital tube display SP2 and the lower row of digital tube display temperature setting value After entering the temperature setting state use the increase and d...

Page 17: ...tton to display TP1 and the lower row shows the first set temperature measurement value Wait for 5 seconds and then click on the decrease view button to exit the first set temperature view state 3 Cir...

Page 18: ...the setting value the cooling 2 compressor has a cooling output Then click the cooling allow button the cooling allow indicator goes out The cooling controller prohibits the cooling output The coolin...

Page 19: ...dicator flashes After the self adjustment is over the indicator light is on and the controller will get a new set of system PID parameters which are automatically saved and the controller automaticall...

Page 20: ...Name Fault Description Sn 1 the fault of the first set temperature sensor or controller itself Sn 2 the fault of the first temperature sensor or controller itself Sn 3 4 20mA current input exceeds th...

Page 21: ...ed for 60 seconds the program will automatically exit this status and the current parameter is not saved Internal parameters table 1 Param eter Parameter Name Parameter Function Description Range Fact...

Page 22: ...eating allow to work 0 1 1 nt nod set value mode Nod 0 set manual adjustment Nod 1 setting value will be adjusted automatically according to the 4 20mA input signal 1 SPL 4 20mA conversion zero positi...

Page 23: ...nal band time proportional action regulation 1 rH 15 I heating integral time integral adjustment function 1 1000 seconds 200 d heating calculus time differential action adjustment 1 1000 seconds 100 P...

Page 24: ...sure to discharge the heat conduction medium XVIII To discharge the heat conduction medium Instruction Connect the verified drain pipeline to the drain outlet and allow the liquid to flow into a suita...

Page 25: ...bar Temperature display setting Touch key input LED digital display resolution is 0 1 degree Safety function Freezer overheat protection sensor abnormal alarm and other overload protections Refrigeran...

Reviews: