LCEN-401P
16
collection tube
Table 12-1 Rotors and adapters
Materials:
Rotor: ABS+PC; Adapters: ABS
3) Notice
⚫
The centrifuge rotor can separate samples with a density lower than 2.0g/ml. If the samples density is
over 2.0g/ml, please calculate allowable speed depending on the following formula.
)
/
density(
ample
)
g/
(
0
.
2
max
)
speed(
llow
ml
g
s
ml
speed
rpm
a
=
4) Autoclaving
A12-10P rotor is made of plastic, cannot be high-pressure sterilization and UV irradiation, only ordinary
sterilization can be used.
12.2 Tubes
1) Cleaning and sterilizing tubes
O
:
Applicable X
:
Inapplicable
Conditions
Materials
PA
PC
PP
C
leani
ng
Cleaning fluids
Acidic
(
pH5 or lower
)
X
X
X
Acidic
(
higher than pH5
)
O
O
O
Alkaline
(
higher than pH9
)
O
X
O
Alkaline
(
pH9 or lower
)
O
O
O
Neutral
(
pH7
)
O
O
O
Warm water(up to 70℃)
O
O
O
Ultrasonic cleaning
Neutral detergent
(
pH7
)
O
O
O
St
er
ili
zat
ion
Autoclaving
115℃
(
0.7kg/cm
2
)
30minutes
O
O
O
121℃
(
1.0kg/cm
2
)
20 minutes
X
O
O
126℃
(
1.4kg/cm
2
)
15 minutes
X
X
X
Boiling
15 to 30 minutes
O
O
O
Ultraviolet sterilization
200-300nm
X
X
X
Gas sterilization
Ethylene oxide
O
X
O
Formaldehyde
O
O
O
PA: Polyallomer PC: Polycarbonate PP: Polypropylene
Table 12-2 Cleaning and sterilizing conditions for tubes
1) Cleaning PC tubes
PC material is low in chemical resistance against alkaline solutions. Avoid using neutral detergents with
pH higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted according
to the manufacturer’s instructions. Use detergent with its pH between 7 and 9.