11. Troubleshooting
11.1 Possible problems and solutions
This centrifuge has a self-diagnostic function. If a problem occurs, an error/warning code will be displayed
on the time display screen, and the operator can determine the malfunction with the alarm code below.
Symptom
Causes
Solutions
Nothing appears on the screen when
the POWER is turned on.
·
Building power circuit breaker
trips.
·
Remove the trouble and turn
on the POWER.
Abnormal vibration
·
Rotor do not match with
spindle
·
Install again the rotor
·
Weighting scales, install
Before lock the rotor, rotate it, and observe carefully if
there is obvious vibration, if so, please take off the
rotor, turn some angle and install it again, until the
rotor rotates smoothly, then, lock it firmly.
Observe here!