Set Cell constant,
this parameter is corresponding to conductivity and salinity
measurement parameters. Usually there are two ways to get Cell constant:
Calibrate the conductive probe with standard conductivity solution, or set the Cell
constant value directly.
Set TDS factor,
this parameter is corresponding to the TDS measurements.
Usually there are two ways to get TDS : Calibrate the conductive probe with
standard conductivity solution, or set the TDS factor directly .
Conductivity compensation,
this parameter is corresponding to conductivity
measurement, the meter supports three kinds of compensation mode, including
no compensation mode, linear compensation mode, pure water compensation
mode.
No compensation: At this point, the measurement reading is absolutely
conductivity value.
Linear compensation mode: The meter will automatically compensate to
the conductivity corresponding to the reference temperature.
Pure water compensation mode: this compensation method is
corresponding to the measurement of pure water.
Conductivity reference temperature,
this parameter is corresponding to the
conductivity measurement. The meter supports two reference temperature: 20.0
and 25.0 , the default is 25.0 .
Salinity compensation mode,
this parameter is corresponding to the salinity
measurement.
Note: This meter does not support seawater mode!
Balance condition,
the meter provides four balance conditions for, as "Fast,
medium, strict, custom", which are corresponding to different parameters and the
measurement environment.
Balance time,
this parameter is corresponding to balance condition, its a
parameter of balance condition, it means the minimum waiting time to meet the
data stability condition. During this time, measurements data should all meet the
balance value, the meter will wait all the time once do not meet. The longer the
time is, the more strict, and vice versa.
Balance value,
this parameter is corresponding to the balance condition, its a
parameter of balance condition. It is the degree of data can be accepted, the
greater the number is, the more easier to balance, and vice versa. Different
measurement parameters have different values.
Note: users can only modify the custom of Balance time, Balance values.
Temperature display, t
he meter supports ( ) Celsius or Fahrenheit ( )
temperature display.
Temperature compensation mode,
each measurement parameter supports
independent temperature compensation, when user selects a manual
temperature compensation, the meter calculate with the manual temperature
value set by the user; When selects automatic compensation, the meter calculate
with the temperature collected from ATC probe. If no ATC probe is connected
with the meter, the meter will still use the manual setting temperature.
Manual temperature,
use manual temperature when automatic temperature
compensation is invalid.