"Calibrate" button to complete the correction.
The meter supports repeating calibrate of the same standard solution, that is
support the function of matching the same solution, if the current correction
solution differ 50uS/cm with previous standard solution, the meter consider
to be the same solution, will automatically cover the calibration data of
previous standard solution. This function can prevent repeated correction, or
allow user to calibrate at will.
If necessary, user will need to reset some correction parameters to
complete correction. Click "Setup" button or sliding around to switch pages.
End correction, press the "<" button to exit the correction, shown in Figure
11.
Calibration Diagram (Figure 11)
2.6.5 Calibration results report
After each end of the calibration, the meter switch to the calibration results pages,
report page indicates the current calibration points, calibration result and the
detailed data of calibration solution.
The meter allows at any time to switch to the calibration results page to view the
calibration results. If there is a previous calibration data, the meter allows to
delete certain calibration results.
At this time, click the "<" button to end correction, shown in Figure 12.