13.3.
Autozero function
The autozero function has been implemented in order to assure precise
indications. This function controls and corrects
„0”
indication. While the
function is active it compares the results continuously with constant
frequency. If two sequentional results differ less than the declared
value of autozero range, so the scale will be automatically zeroed
and the pictograms
and
will be displayed.
When AUTOZERO is disabled zero is not corrected automatically.
However, in particular cases, this function can disrupt the measurement
process e.g. slow pouring of liquid or powder on the weighing pan.
In this case, it is advisable to disable the autozero function.
Procedure:
•
Enter the submenu
<P1.rEAd>
and then:
Fnnd
no
- filter disabled
Fnnd
YES
- filter enabled
Return to weighing:
See - 11.2.2.
- 27 -