Then the program automatically goes to the internal calibration
procedure which is signalled by the following message:
−
After completion of the calibration process program returns
to the weighing mode.
•
Calibration triggered off by timer
−
The time condition for subsequent automatic calibration is 3
hours. It means that, when no other triggering factor appear,
the calibration will appear every 3 hours;
−
The calibration procedure triggered off by the time change starts
with checking which is signalled by the following message:
−
Then the program automatically goes to the internal calibration
procedure which is signalled by the following message:
−
The program for non-verified scales has a parameter for setting
a maximal time interval between subsequent internal calibration.
Procedure:
−
Enter the submenu
<P6.CAL>
and then:
- 60 -