Using
and
set
the tare value
,
•
Press
,
•
Program returns to weighings mode. The inscribed tare value
can be seen on the display with
„–”
sign,
•
Tare can be inscribed anytime in weighings mode.
Notice:
1.
You cannot inscribe a new tare value when the tare value in
memory
is greater than zero. In the case of trying this the
<Err3>
message
will be displayed and short audible signal will be emitted.
2.
Users can also enter up to 9 tare values to the scale
memory
(see 16.10 of his manual).
12.3.
Zeroing
To
ZERO
the scale press:
.
The scale will display zero and following pictograms:
and
.
Zeroing is only possible within the scope of
±2%
of full scale. While
zeroing outside the scope of
±2%
you will see
<Err2>
. Zeroing
is possible only in stable state.
Notice:
Zeroing is possible only within the
±2%
interval of the maximal range.
If zeroing is performed beyond this range the
<Err2>
message and
short audible signal will be emitted.
12.4.
Weighings in two ranges
Switching between the
I range
and the
II range
happens automatically
(exceeding Max of the
I range
). Weighings in the second range is
signalled by a pictogram in the top left corner of the display.
- 21 -
Summary of Contents for BWLC-A1-C2
Page 2: ...www mrclab com 2 ...
Page 3: ...MARCH 2010 3 ...
Page 11: ...Scales BWLC C1 R Scales BWLC C2 R Scales BWLC C1 K Scales BWLC C2 K 11 ...
Page 31: ...Return to weighing see 11 2 2 14 3 Baud rate Procedure Enter the submenu P2 Prnt and then 31 ...
Page 46: ...but deviation of load mass placed on the pan referring to the mass of standard in 46 ...