The completion of the calibration procedure is signalled
by the following message:
•
Take off the weight from the pan, the message
<donE>
is
displayed for 1s and the program returns to the calibration submenu:
•
Calibration process can be terminated anytime by pressing
which is signalled by the following message on the display:
•
Return to weighing with saving changes that have been
made.
Caution:
If the calibration process (span adjustment) lasts longer than 15 the
<Err8>
message will be displayed and short audible signal will be
emitted. Press
to perform calibration again with more stable
ambient conditions!
17.3.
Start mass adjustment
Option only for non-verified scales
Not accessible in
B
WLC…C2 scales of
B
WLC series
The scale can require only adjusting new start mass. In this scale adjusting
start mass can be excluded from the process of calibration and performed
separately.