Online Operation
3 Online Operation
1. Insert
WASTE FLUID HOSE
in to the waste disposal bottle pro-
vided.
Important! Do not connect Waste Fluid Hose to a pressurised
system, as this will cause the Analyser to malfunction and could
cause internal leakage. The Waste Fluid Hose must be dis-
charged into the waste disposal bottle provided, or into a tank/ves-
sel vented to atmosphere.
2. Connect
WASTE FLUID HOSE
to Analyser (waste connector). Push
back quick coupling outer ring before connecting / disconnecting
hose end.
3. Connect
FLUID SAMPLING HOSE
to Analyser (HP connector)
4. Connect
FLUID SAMPLING HOSE
to the system by means of the
minimess connector.
The system to be monitored must not exceed 400 bar or be less
than 2 bar
5. Press
GREEN BUTTON
to switch on Analyser the "Main / test
progress screen’’ will be displayed.
To prolong battery life it is advisable to switch off the Analyser
when not in use.