Continuous Sampling
21
5 used to input the Clean Alarm Level when using the NAS1638
/ AS4059E-1 format. Enter 0 if not used.
6 used to enter the Clean Alarm Level when using the AS4059E-2
format. Enter 0 if not used. (This Alarm Level will now be
displayed as *A/*B/*C/*D/*E/*F).
3
Press the flush valve push button to open the flush valve. The push
button illuminates to indicate that the valve is open. Leave valve
open for at least 1 minute or 200ml of fluid, or more if the HP sam-
pling hose is greater than 1.5m long.
4
Press flush valve push button to close flush valve – push button il-
luminates is cancelled. Alternatively, proceed to Step 5 below – the
action of pressing Start button automatically closes the flush valve
before sampling commenced.
5
Press the
START
button 1
The LPA² will now commence the sampling procedure.
6
The completion progress bar indicates the status of the test.
•
Results will be automatically displayed on the screen after each
test.
•
Results will be automatically printed at the end of the emptying
cycle, if the Auto Print mode has been turned
ON
7
The status is shown as Waiting between the ending of one test and
the starting of the next
8
Press Stop button 2 at any point in the cycle to end continuous
sampling. The test status will show Idle.