7 Troubleshooting / FAQ
7.1 Misuse of Product
- The product should be connected to a power supply within the rating of the product and not wired directly to the mains.
- Ensure that the correct version of the product is being used with the correct fluids, not all versions are compatible with all fluids.
- The operator should follow all standard operating procedures previously set at the operating location as well as the procedures
required by the manufacturer.
- The Bottle Sampler is not suitable for use in an explosive environment or an ATEX zone.
TROUBLESHOOTING / FAQ
7.2 Fault Finding
Fault
Check
Unexpected results obtained from sample
Ensure that the microbore pressure hose has been fully connected at both the
bottle sampler and the CMP. Confirm that there is a free flow of fluid to the CMP,
by operating the flush valve and observing fluid passing to waste.
Check that the BS500 is reaching sufficient pressure, 2 to 2.5 bar.
Bottle Sampler not reaching required pressure
Regrease O-ring in the base of bottle sampler using a
suitable grease, silicone grease is recommended.
Sporadic results -
especially when using triple & bottle testing
The bottle sampler has reached pressure before starting the test.
Also that the bottle sampler is powered for full duration of the test.
Fluid level is not changing during flush/test
The bottle sampler has reached full pressure before starting test/flush.
Viscosity of fluid is below 400cSt
BS110 - the dip tube has been inserted and is still attached.
The Microbore hoses are fully engaged at both ends.
Operators Guide
31
Summary of Contents for BS110
Page 1: ...Operators Guide BS110 BS500 Bottle Samplers ...
Page 2: ......
Page 6: ...Operators Guide 4 ...
Page 8: ...Operators Guide 6 ...
Page 11: ...OPERATOR S GUIDE Operators Guide 9 ...
Page 22: ...Operators Guide 20 ...
Page 32: ...Operators Guide 30 ...
Page 35: ......