In case Default sample tube holder is to be changed, then press SET key, LCD will
show 4.0 M/S , now use
p
key to reach upto the message MP:24*2. Press SET key
when this is displayed, LCD will start Flashing, to indicate the sample tube holder can
be changed using
p
or
q
key, Once tube holder is selected, press SET key again
,Flashing of LCD will stop to indicate that tube holder has been selected.
NOTE: During tube holder selection LCD will be displayed as
MP:24*2 - QuickPrep Adapter,( Supplied with instrument)
BG:2*50 - BigPrep Adapter, (2 tubes of 50 ml)
TN:12*15 -TeenPrep Adapter, (12 tubes of 15 ml)
CY:24*2 -CryoPrep Adapter, (24tubes of 2 ml)
HG:48*2 - HighPrep Adapter, (48 tubes of 2ml)
CUSTOM - Custom holder
In case Default Time is to be changed, then Press SET key, LCD will show
4.0 M/S
now use
p
key to reach upto the message
TIME 20
, press SET key when this is
displayed, LCD will start Flashing, to indicate the Run time can be changed using
p
or
q
keys.
Once Run time is selected, press SET key again,
Flashing of LCD will stop
to indicate that Run time has been selected.
Once speed, tube holder and time have been selected or saved, Press RUN key,
LCD will show
SURE ?
, press RUN key to start the session.
If you are ready to start, then press the RUN Key again. The motor will start and
display will show count down run time.
If you are not sure then press SET key for selection and changing of desired
speed, tube holder and time.
When the time is over, LCD will display
REST 5 Mn
indicating that the motor is at rest for 5 minutes, display will count down to 4, 3, 2, 1
minute. LCD will display
READY
, AND the program will go back to main menu with
last selection.
Please note, when Speed, tube holder or Run Time Value is flashing [in Selectable
Mode], than during this user can still start the unit directly by pressing
RUN
key
twice, but these values of Speed or Time will be temporary and they will not be saved
unless SET key is pressed and LCD stops flashing.
13
www.mpbio.com
TM
TM
TM
TM
TM