background image

 

 

TABLE OF CONTENTS 

 

SECTION 

 

II

 

Specifications ................................................................................................................ 4

 

III

 

Setting-up the Instrument .............................................................................................. 5

 

IV

 

Assembling the microscope Input voltage ..................................................................... 6

 

1

 

Installing the lamp .................................................................................................................... 6

 

2

 

Mounting the condenser ........................................................................................................... 6

 

3

 

Installing Objectives ................................................................................................................. 7

 

4

 

Mounting the eyepieces ........................................................................................................... 7

 

 

V

 

Microscopic procedure Interpupillary distance adjustment ............................................ 7

 

1

 

Diopter adjustment ................................................................................................................... 8

 

2

 

Centering the Condenser ......................................................................................................... 8

 

3

 

Centering the lamp ................................................................................................................... 9

 

4

 

Brightfield microscopy .............................................................................................................11

 

5

 

Phase-contrast microscopy .....................................................................................................11

 

 

VI

 

Photo Procedure .......................................................................................................... 13

 

VII

 

Troubleshooting Table ................................................................................................. 15

 

VIII

 

Care and maintenance ................................................................................................ 16

 

1

 

Lenses and filters ....................................................................................................................16

 

2

 

Cleaning of painted or plastic components ..............................................................................16

 

3

 

When not in use ......................................................................................................................16

 

 

IX

 

Warranty ...................................................................................................................... 17

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for SWIFTLINE MAE-31R

Page 1: ...1 Inverted Microscope Instruction Manual Model MAE 31R...

Page 2: ...ounting the eyepieces 7 V Microscopic procedure Interpupillary distance adjustment 7 1 Diopter adjustment 8 2 Centering the Condenser 8 3 Centering the lamp 9 4 Brightfield microscopy 11 5 Phase contr...

Page 3: ...Clamp Screw 9 Revolving Nosepiece 3 Annular Diaphragm 10 Coaxial Coarse Fine Focus Knob 4 Condenser Diaphragm Lever 11 Torque Adjusting Ring 5 Condenser Focus Knob 12 Filter Retaining Ring 6 Condenser...

Page 4: ...r 6 Filter Slider 7 Lamp House 8 Field Diaphragm Lever 9 Condenser Clamp Screw 10 Phase Slider 11 Condenser Clamp Holder Screw 12 Condenser Centering Screws 13 Stage Plate 14 Fluorescence Filter casse...

Page 5: ...on N A W D mm Plan Achromat 4X 0 10 23 5 Plan Achromat Phase 10X 0 25 7 5 Plan Achromat Phase LWD 20X 0 40 7 0 Condenser Extra Long Working Distance N A 0 3 W D 72mm Electrical Specifications Input IE...

Page 6: ...y damage the electrical components of the microscope In order to prevent electric shock always turn the power switch on the power supply off before connecting the power cord 1 Installing the lamp In o...

Page 7: ...into the nosepiece so that the magnification increases with clockwise rotation of the revolving nosepiece Replace the stage plate insert 4 Mounting the eyepieces Remove the dust caps from the eyepiece...

Page 8: ...ocus using the 10x objective Turn the diopter compensation ring on each eyepiece until the adjustment ring is adjusted to 0 position Position 40x objective into the optical path and bring the specimen...

Page 9: ...f view Adjust the condenser centering screws so that the center of the field diaphragm image matches the center of the field of view This adjustment is easier to make if the field diaphragm size is re...

Page 10: ...low the knob to move around with little resistance Holding the lamp socket clamp screw move the lamp filament image to the center of the condenser diaphragm iris The filament will be visible looking d...

Page 11: ...t is important because it determines the resolution of the image contrast depth of focus and brightness Closing down the aperture diaphragm will lower the resolution and brightness but increase the co...

Page 12: ...be and insert the phase centering telescope in its place Loosen the locking screw of the centering telescope eyepiece and slide out until both the phase plate image of the objective and the annular di...

Page 13: ...e filter in the designated retaining ring above the phase annular diaphragm slider Filter selection Filter type Procedure GIF Green interference 546nm For phase contrast and contrast adjustment with b...

Page 14: ...mera simply screw onto your C Mount camera Remove the cap on the camera port of your microscope by loosening the knurled screw Then attach the C0 5X and lock into place with the knurled screw Then fol...

Page 15: ...tfield objective being used Phase annular diaphragm not in optical path Phase annular diaphragm and objective phase symbol do not match Slider annular ring image has moved away from the objective phas...

Page 16: ...enzine and absolute alcohol are both highly flammable be careful handling around open flame Do not use same area of gauze or tissue to wipe more than once 2 Cleaning of painted or plastic components D...

Page 17: ...ceipt with the RMA number also clearly written on the exterior of the shipping box Microscope s must be shipped back to originating location or sales office A detailed description of the difficulty mu...

Reviews: