background image

 

15 

 

VII  Troubleshooting Table 

 

The troubleshooting table below contains the most frequently encountered problems and their possible 

causes. 

 

Optical and Operating Problems 

 

Problem 

Possible Cause 

Vignetting  or  uneven  brightness 
in the field of view or field of view 
only partially visible 
 
 

Lamp not installed properly 
Filter slider in intermediate position 
Phase slider not in click-stop position 
Incorrect condenser mounting 
Condenser is set too low 
Condenser is not centered 
Field diaphragm closed too far 
Aperture diaphragm closed too far 
Revolving nosepiece not clicked   
into position 
Optical path selector lever in   
intermediate position 

Dust or dirt in field of view 

Aperture diaphragm closed too far 
Field of view diaphragm image not   
focused on specimen surface 
Dust or dirt on specimen’s surface 

Image quality: 
No  image  under  phase  contrast 
or details cannot be viewed   

Brightfield objective being used 
Phase annular diaphragm not in   
optical path 
Phase annular diaphragm and   
objective phase symbol do not match 
Slider annular ring image has   
moved away from the objective   
phase plate image 
Field of view diaphragm image not   
focused on specimen surface 
Thickness of specimen holder is   
outside the compensating range   
of objective 

Eye strain or fatigue 

Interpupillary distance not adjusted 
Diopter adjustment not made 
Inadequate illumination 
Field of view of left and right eyepiece differ 

 

Electrical 

 

Lamp does not light 

Power supply not plugged in 
Lamp not installed 
Lamp burnt out 

– 100hrs Average Life 

Inadequate brightness 

Specified lamp not being used 

Lamp blows out immediately 

Specified lamp not being used 

Lamp flickers 

Connectors are not securely connected 
Lamp near end of service life 
Lamp not securely plugged into socket 

   

Never attempt either of the following actions, since doing so will damage 

the focusing mechanism: 

o

 

Rotate the left and right knob while holding the other. 

o

 

Turning the coarse and fine focus knobs further than their limit. 

 

Summary of Contents for SWIFTLINE MAE-31R

Page 1: ...1 Inverted Microscope Instruction Manual Model MAE 31R...

Page 2: ...ounting the eyepieces 7 V Microscopic procedure Interpupillary distance adjustment 7 1 Diopter adjustment 8 2 Centering the Condenser 8 3 Centering the lamp 9 4 Brightfield microscopy 11 5 Phase contr...

Page 3: ...Clamp Screw 9 Revolving Nosepiece 3 Annular Diaphragm 10 Coaxial Coarse Fine Focus Knob 4 Condenser Diaphragm Lever 11 Torque Adjusting Ring 5 Condenser Focus Knob 12 Filter Retaining Ring 6 Condenser...

Page 4: ...r 6 Filter Slider 7 Lamp House 8 Field Diaphragm Lever 9 Condenser Clamp Screw 10 Phase Slider 11 Condenser Clamp Holder Screw 12 Condenser Centering Screws 13 Stage Plate 14 Fluorescence Filter casse...

Page 5: ...on N A W D mm Plan Achromat 4X 0 10 23 5 Plan Achromat Phase 10X 0 25 7 5 Plan Achromat Phase LWD 20X 0 40 7 0 Condenser Extra Long Working Distance N A 0 3 W D 72mm Electrical Specifications Input IE...

Page 6: ...y damage the electrical components of the microscope In order to prevent electric shock always turn the power switch on the power supply off before connecting the power cord 1 Installing the lamp In o...

Page 7: ...into the nosepiece so that the magnification increases with clockwise rotation of the revolving nosepiece Replace the stage plate insert 4 Mounting the eyepieces Remove the dust caps from the eyepiece...

Page 8: ...ocus using the 10x objective Turn the diopter compensation ring on each eyepiece until the adjustment ring is adjusted to 0 position Position 40x objective into the optical path and bring the specimen...

Page 9: ...f view Adjust the condenser centering screws so that the center of the field diaphragm image matches the center of the field of view This adjustment is easier to make if the field diaphragm size is re...

Page 10: ...low the knob to move around with little resistance Holding the lamp socket clamp screw move the lamp filament image to the center of the condenser diaphragm iris The filament will be visible looking d...

Page 11: ...t is important because it determines the resolution of the image contrast depth of focus and brightness Closing down the aperture diaphragm will lower the resolution and brightness but increase the co...

Page 12: ...be and insert the phase centering telescope in its place Loosen the locking screw of the centering telescope eyepiece and slide out until both the phase plate image of the objective and the annular di...

Page 13: ...e filter in the designated retaining ring above the phase annular diaphragm slider Filter selection Filter type Procedure GIF Green interference 546nm For phase contrast and contrast adjustment with b...

Page 14: ...mera simply screw onto your C Mount camera Remove the cap on the camera port of your microscope by loosening the knurled screw Then attach the C0 5X and lock into place with the knurled screw Then fol...

Page 15: ...tfield objective being used Phase annular diaphragm not in optical path Phase annular diaphragm and objective phase symbol do not match Slider annular ring image has moved away from the objective phas...

Page 16: ...enzine and absolute alcohol are both highly flammable be careful handling around open flame Do not use same area of gauze or tissue to wipe more than once 2 Cleaning of painted or plastic components D...

Page 17: ...ceipt with the RMA number also clearly written on the exterior of the shipping box Microscope s must be shipped back to originating location or sales office A detailed description of the difficulty mu...

Reviews: