For differences in eyesight: 

 

1.  Using only the right eye, look through the right eyepiece (1) and adjust focus. 

 

2.  Next,  using  the  left  eye,  look  through  the  eyepiece  and  adjust  the  focus  by  turning  the  diopter 

adjuster (2) located on the left-hand tube (3) until the image appears sharp. Do not use the focus 
knobs to adjust focus (18). 

 

E.  Changing magnification. 

 

1.  Turn the zoom control (16) to the highest magnification, 4X. 

 

2.  Although  the  stereomicroscope  has  been  parfocalised,  focus  will  have  to  be  adjusted  as  the 

objectives of a low magnification offer a more profound field of view. The profundity of the field is 
the capacity to focus on different points on different levels. 

  

3.  Once the image is in focus with the higher magnification objectives it is not necessary to adjust 

the focus when lower magnification objectives are used.  

 

 
Adapting a photographic, or a video camera 

 (Only for model SMZ-143) 

 
The  model  SMZ-143  comes  equipped  with  a  vertical  image 
port  on  the  head  for  the  installation  of  a  photographic  reflex 
type  camera,  or  video  camera,  using  the  corresponding 
adapters.  
 
At the back of the head there is a selection lever (Fig. 2) that 
transmits the image to the vertical port. In an extended position 
the  image  is  transmitted  to  the  vertical  port,  but  the  image 
cannot be seen by the right eyepiece.  
 
A. 

To  adapt  a  photographic  camera  an  adapter  tube  is 
needed.  This  tube  includes  2X  lenses  to  balance  the 
correct parfocality between the images received by the 
binocular  and  vertical  ports.  This  balance  can  be 
achieved  providing  that  the  parfocality  has  been 
adjusted  correctly,  as  described  in  point  “D.  Changing 
magnification” in this manual.    

 
 

The adapter tube has a T type thread at one end where adapter mounts of all types can be adapted 
for all reflex camera brand names on the market.   

 

NB

: The T type adapter mount is not included as it is an item specific to each brand of camera on the 

market 

 

1.  To connect the camera to the microscope, first remove the front lens of the camera, and place the 

corresponding T mount. Screw on the adapter tube provided, onto the mount of the camera.  

 

2.  Loosen the knurled screw (Fig.2), situated on the side of the vertical port on the head, sufficiently 

enough to remove the protective cover.  

 

3.  Insert the adapter tube with the camera already mounted on the vertical port. If it does not insert 

easily, unscrew the knurled screw until the adapter tube fits perfectly. 

 

4.  Re-tighten the knurled screw, so that the camera is secure. 

 

5.  Operate the camera according to manufacturers instructions. 

 

-6- 

 

Selection 

lever 

Fixing 

screws 

Adapter 

tube 

Fig. 2 

Summary of Contents for SMZ Series

Page 1: ...Instruction Manual SMZ Series Copyright 02 02 Motic Microscopes European Division...

Page 2: ...Eyepiece 14 Head 18 Focus knob 20 Support collar screw 22 Incident illumination switch I 23 Transmitted illumination switch T 24 Light intensity control 2 Diopter adjuster 3 Eyepiece tubes 5 Head hold...

Page 3: ...ents of the stereomicroscope have been carefully packed to make sure they reach you in perfect condition We recommend that you do not discard any packing containers in case you need to return the micr...

Page 4: ...ent objects or for dissection For best contrast choose the side of the stage to use Warning Transmitted illumination can ONLY be used with the frosted glass stage THE HEAT GENERATED BY THE TRANSMITTED...

Page 5: ...ntre of the stage 11 3 Turn focussing knobs 18 to mid focus range 4 The head holder 17 is mounted on a column 21 on which it can be moved up or down depending on the size of the object to be focussed...

Page 6: ...xtended position the image is transmitted to the vertical port but the image cannot be seen by the right eyepiece A To adapt a photographic camera an adapter tube is needed This tube includes 2X lense...

Page 7: ...hat the CS mount is responsible Place or remove it according to the procedure to obtain parfocality Maintenance WARNING FOR YOUR OWN SAFETY SWITCH OFF AND DISCONNECT THE MICROSCOPE FROM ANY ELECTRICAL...

Page 8: ...ination bulb a Unscrew the illuminator protector tube 7 turning it anti clockwise and removing the tube from that of the lenses b With a cloth carefully pull out the bulb to disconnect it from the soc...

Page 9: ...ly as possible without the head sliding down with its own weight The tension adjustment collar for focussing Fig 1 is situated between the focussing knob 18 and the head holder 17 a Loosen the screw l...

Page 10: ...use cover badly closed Bad connection with power source Insert bulb correctly Replace bulb Close correctly Have repaired by a qualified specialised technician Fuse blown quickly Wrong fuse Replace wit...

Page 11: ...ils of the required revision Warrantee All MOTIC microscopes are warranted against any manufacturing defect for a 5 year period Damage occurring by any unauthorised repair work or occurring through mi...