SECTION 4
32
Configuration Page 5
Plot Format
Defines the default for one of the available plot formats in either standard or Cabrera presentation. Please note
that regardless of the plot format selected, 10 seconds of 12 leads are always stored.
The ECG plot options are:
Rhythm Leads
Identifies the three user-selectable, 10-second rhythm leads for the 3+1 and 3+3 channel ECG printout.
NOTE:
Rhythm acquisition is not stored in memory, only printed.
NOTE:
See Section 3 to acquire a rhythm printout.
Bar Code Scanner
Allows a bar code scanner to be connected to the serial port on the rear of the device to allow scanned input
of patient demographics information printed on the patient’s wristband. The bar code scanner can also
function as a keyboard wedge and input certain function commands such as AUTO, RHY, XMT, and STOP.
Average RR
Enabling this option will display an averaged RR value to appear on the report.
QTcB
Enabling this option will display a Bazett’s corrected QT value on the report along with the default linear
QTc value.
QTcF
Enabling this option will display a Fridericia corrected QT value on the report along with the default linear
QTc value.
Format Option
ECG Data
2.5 seconds of 12 leads in a 3-channel format and
3+1
the fourth channel is a 10-second rhythm strip of
Lead II (user defined). Cabrera also available.
3
2.5 seconds of 12 leads in a 3-channel format.
Cabrera also available.
2.5 seconds of 12 leads in a 3-channel format plus
3+3
10-second rhythm strip of Leads II, V1, and V5
(user defined) in a 3-channel format. Cabrera
also available.
Summary of Contents for ELI 150 Rx
Page 12: ...User Safety Information Continued x...
Page 25: ...ELI 150 Rx Left Side Figure 1 2 ELI 150 Rx Rear Figure 1 3 SECTION 1 3...
Page 26: ...SECTION 1 ELI 150 Rx Base Figure 1 4 4 T5A 250V...
Page 34: ...SECTION 2 12...
Page 44: ...SECTION 3 22...
Page 58: ...SECTION 4 36...
Page 74: ...APPENDIX B 52...