Operation 2011-07-05
33
*1: Charger Data
*2: This test is not applicable since the EUT signal cable is less than 3 m.
Appendix- Electromagnetic declaration
Guidance and manufacturer’s declaration – electromagnetic emissions
The TR-CM is intended for use in the electromagnetic environment specified below. The customer or the user of the TR-CM should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The TR-CM uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The TR-CM is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
*1
IEC61000-3-2
Class A
Voltage
fluctuations/ flicker emissions
*1
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The TR-CM is intended for use in the electromagnetic environment specified below. The customer or the user of the TR-CM should assure
that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
+6 kV contact +8 kV air
+2, 4, 6 kV contact +2, 4,
8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transients/
bursts
IEC 61000-4-4
+2 kV for power supply
lines +1 kV for input/
output lines
+2.0 kV for power supply
lines
*2
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
+1 kV line(s) to line(s) +2
kV line(s) to earth
+0.5, 1, 2 kV line(s) to
line(s) +0.5, 1 kV line(s)
to earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply lines IEC 61000-
4-11
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in UT )
for 5 cycles
70% UT (30% dip in UT )
for 25 cycles
<5% UT (>95% dip in
UT) for 5 sec
0% UT (>95% dip in UT)
/0.5 periods
40% UT (60% dip in UT )
/5 periods
70% UT (30% dip in UT )
/25 periods
0% UT /5 sec.
Mains power quality should be that of a typical
commercial or hospital environment. If user of the
TR-CM requires continued operation during power
mains interruptions, it is recommended that the
TR-CM be powered from an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3.15 A/m
Power frequency magnetic field should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.