Morita Tri auto mini Setup Instructions Download Page 21

Operation   2011-07-05

20

Meter Display

Operation

1

iii

ii

i

Measurement Bar

Meter Numbers

Flash Bar

0.5

Blinks On and Off

• 

Occasionally the meter will make a sudden and large movement as soon as the file is inserted into the root canal, but it will return to normal as the file is 

advanced down towards the apex.

• 

The contrary electrode, file electrode and metal parts of the contra angle could cause an adverse reaction if the patient has an allergy to metals.  Ask the 

patient about this before using the Tri Auto mini.

•  Take care that medicinal solutions such as formalin cresol (FC) or sodium hypochlorite do not get on the contrary electrode or the contra angle.  These could 

cause an adverse reaction such as inflammation.

• 

The file electrode cannot be used with the following types of files including nickel-titanium ones.  Use these files without attaching the file electrode.

  Files with a shank diameter greater than 1.2 mm, Files with shanks that do not have a circular cross section, Gates Glidden Drills, Tools with large cutting 

heads such as largo burrs.

•  In some cases such as a blocked root canal, a measurement can not be made.  (For details refer to the section of the Root ZX mini manual that covers canals 

not suitable for measurement.)

•  Accurate measurement is not always possible, especially in cases of abnormal or unusual root canal morphology; always take an X-ray to check the 

measurement results.

• 

If the meter does not move when the file is inserted, the unit may be malfunctioning and must not be used.

•  Do not use an ultra sonic scaler while the contrary electrode is hooked in the patient’s mouth; noise from the scaler could cause the motor to start running 

resulting in an accident or injury.

• 

Absolutely never allow the contrary electrode, the handpiece file electrode or the connections for these to contact an ordinary AC power source such as a 

socket; this could result in a very serious and angerous shock.

* Refer to the user manual for the Root ZX mini for information about canal measurement and for warnings and notes about use.

i.  The Measurement Bar shows the location of the file tip.  The Flash Bar blinks on and off when the file is inside the canal.  
ii.  The 0.5 meter reading shows where the file tip is about 0.5 to 1.0 mm from the anatomical apex.

* The numbers 1, 2, and 3 on the meter do not indicate length in millimeters.

iii.  If the file tip goes past the Flash Bar, an alarm will sound and the backlight will blink on and off.

Contrary Electrode

Turn on the Tri Auto mini and Root ZX mini.
The backlight for the display will be yellow.
Hook the Contrary Electrode in the corner of the 
patient’s mouth.

Summary of Contents for Tri auto mini

Page 1: ...Cordless endodontic treatment motorized handpiece Tri Auto mini TR CM Operation Instructions...

Page 2: ...2 Operation 12 Basic Operation 12 Memory Settings 13 Memory Settings Primary Functions 13 Memory Settings Additional Operation Settings 14 Memory Settings Other Settings 15 Restore Default Memories 16...

Page 3: ...N 1394779 CN DESIGN APPLN 201030166890 6 Thank you for purchasing the Tri Auto mini For optimum safety and performance read this manual thoroughly before using the unit and pay close attention to warn...

Page 4: ...oided by foreseeing the possibility of danger and operating the unit in accordance with the manufacturer s recommendations First thoroughly read all precautions and instructions pertaining to safety a...

Page 5: ...RITA MFG CORP and other than in their original condition 5 operating the equipment in ways other than the operating procedures described in this manual or resulting from the safety precautions and war...

Page 6: ...ce electrical noise Do not use this instrument on patients who have a pacemaker or an Implantable Cardioverter Defibrillator ICD It could cause the pacemaker or the Implantable Cardioverter Defibrilla...

Page 7: ...automatically reverses its rotation if the torque load exceeds the set value to reduce the risk of jamming Slow Down The file slows down as torque increases The file slows down as it approaches the a...

Page 8: ...lus and Minus Switches Power Supply Cord Charge LED orange Power LED green Charger Motor Handpiece Charger Switch Parts Identification and Accessories Battery Inside motor handpiece Transmission Cable...

Page 9: ...ight get wet Usage Operation Conditions for the main unit and AC adapter Charge the battery as soon as the battery power indicator gets down to its last bar If the orange charge LED goes off immediate...

Page 10: ...s frequently Contra Angle Motor Handpiece Line up the marks Approx 290 Do not force the contra angle to rotate beyond its limit The rotation range of the contra angle is about 290 Set it so that it re...

Page 11: ...cted to the Root ZX mini Never use deformed or damaged files Give the file a light tug to confirm it is securely held in place If the file is not securely placed it could come out and injure the patie...

Page 12: ...y to check Nickel titanium file may suddenly snap depending on the curvature and shape of the canal stop using the file if you notice or feel anything amiss Files will eventually break due to metal fa...

Page 13: ...motor Press Main switch again The Torque Display will appear If you hold the Main Switch down when you start the motor it will run only while the switch is held down and stop when the switch is relea...

Page 14: ...0 1 5 2 0 2 5 and 3 0 Ncm This function can also be turned off TRL torque reverse less Rotation Direction Fwd Forward Rev Reverse Standby Display Rotation Direction Speed Torque Reverse These functio...

Page 15: ...Plus or Minus switch The display will go back to the standby display if 5 seconds factory setting elapses without a switch being pressed Torque Slow Down When this is turned on the motor will slow dow...

Page 16: ...Change Return to Standby Time Beeper Volume Press Plus or Minus switch to set beep volume used for switch operation and alarms at Off Low or Big Auto Power Off Time The time lapse for automatic shut...

Page 17: ...s or press the Main switch to cancel the operation 3 Wait unit the Finished display appears and then press the Main switch to go to the standby display Take care not to injure your fingers when insert...

Page 18: ...used to secure the file electrode Take out the black screw and remove the cap that covers the elctrode installation area Usage Operation with the Root ZX mini Electrode Site Cap File Electrode Instal...

Page 19: ...s without holding the file release button down will damage the chuck Use caution when inserting and removing files to avoid injury to fingers Make sure the file goes all the way in Give the file a lig...

Page 20: ...up when you do this Make sure the plugs go straight in After insertion give plugs a light tug to make sure they are securely connected Otherwise data may not be transmitted acurately Do not bump the...

Page 21: ...not suitable for measurement Accurate measurement is not always possible especially in cases of abnormal or unusual root canal morphology always take an X ray to check the measurement results If the...

Page 22: ...single sustained beep will sound when this happens If the unit is set for Apical Reverse the motor will run backwards after it stops If the load on the file exceeds the setting for torque reverse the...

Page 23: ...er goes all the way to its end and there are no segments that do not light up Refer to page 17 for how to install the new file electrode Take out the screw Take out the old file electrode Pull the fil...

Page 24: ...r each patient Recommended 134 C 273 2 F for at least 5 minutes inside a sterilization pouch or 121 C 249 8 F for at least 35 minutes inside a sterilization pouch Minimum drying time after sterilizati...

Page 25: ...ther components These chemicals could damage deform or discolor plastic and metal Use extra caution to avoid spilling formalin cresol FC and sodium hypochlorite as they are quite strong Wipe up any ch...

Page 26: ...2 for how to replace the file electrode Connect the new battery and put it in Replace the cover and its screw Turn power off Remove the screw and slide the battery cover off Take out the old battery a...

Page 27: ...45 C 14 to 113 F Realtive Humidity 10 to 85 RH without condensation Atmospheric Pressure 700 to 1 060 hPa Do not expose to direct sunlight frequently or for long times If the unit has not been used fo...

Page 28: ...pressed while the Select switch is being held down 3 Check that pressing the Plus and Minus switches changes the memory number from M1 through M6 4 Check that the settings for each memory can be chan...

Page 29: ...ld be recycled Metal parts of the equipment are disposed as scrap metal Synthetic materials electrical components and printed circuit boards are disposed as electrical scrap Material must be disposed...

Page 30: ...setting turned off Turn Linked Function setting on Is Auto Start Stop turned off Turn Auto Start Stop setting on Has the meter gone past the Flash Bar Set the Apical Stop or Reverse for Reverse REV M...

Page 31: ...oard may be defective Micromotor changes speed on its own Is Apical Slow Down setting tuned on When this is turned on the motor slows down as the file approaches the apex Refer to page 14 for how to t...

Page 32: ...itting Length of Shank 9 0 mm Maximum Overall Length of Rotary Instrument 28 0 mm Type of Shank Type 1 according to ISO 1797 1 Type of Coupling TR CM coupling It can be used only with TR CM motor hand...

Page 33: ...equipment cannot be disposed of as unsorted municipal waste within the European Union Follow local regulations for disposal Attention consult accompanying documents Manufacturer Type BF applied part C...

Page 34: ...nt guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 2 4 6 kV contact 2 4 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material t...

Page 35: ...hould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency...

Page 36: ...ications equipment transmitters and the TR CM as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance...

Page 37: ...wn should be devised to protect the safety of the patient in case any abnormalities in the equipment or the patient are observed 4 Make sure the patient does not handle or manipulate the equipment 5 I...

Page 38: ......

Page 39: ......

Page 40: ...PUB M9826 EN 4 Printed in Japan...

Reviews: