14
Torque Slow Down
Linked Function
Auto Start & Stop
Apical Slow Down
Apical Reverse or Stop
Apical Torque Reduction
Memory Settings: Additional Operation Settings
Additional Functions
:
Torque Slow Down (TORQ.SL.D.), Linked to canal measurement (LINKED FUNCTION
**
),
Apical Reverse or Stop (APICAL ACT.
**
), Auto Start and Stop (AUTO STT/STP
**
), Apical Slow Down
(APICAL SL.D.
**
), Apical Torque Reduction (APICAL TRQ.D.
**
)
1. Select a memory number for the standby display; press the Plus or Minus switch.
2. Hold down the Select switch for at least 1 second to show the displays for
additional operation settings.
3. Press the Select switch to go from one display to the next.
4. Change the setting; press the Plus or Minus switch.
* The display will go back to the standby display if 5 seconds (factory setting) elapses without a
switch being pressed.
Torque Slow Down:
When this is turned on, the motor will slow down as the torque load increases.
Standby Display
Linked Function
**
:
When this is turned on, the functions below will be activated.
Apical Reverse or Stop
**
:
The fil
e will reverse o
r stop when the file tip reaches the Flash Bar.
Auto Start & Stop
**
:
When this is turned on, the file starts rotating when it is inserted and stops when it is taken out
of the canal.
Apical Slow Down
**
:
When this is turned on, the
file slows down as it approaches the apex.
Apical Torque Reduction
**
:
When this is turned on, the torque setting that triggers reverse rotation is reduced as the file
tip approaches the apex.
**
These functions are available only when connected to the Root ZX mini.