27
Regular Inspection
Maintenance and Inspection Items
Parts List
Maintenance and Inspection
* Maintenance and inspection are generally consider to be the duty and obligation of the user, but if, for some reason, the user
is unable to carry out these duties, he may rely on a qualified medical device serviceman. Contact your local dealer or J.
MORITA OFFICE for details.
* Replace the parts listed in the Parts Lists as necessary depending on degree of wear and length of use.
* This apparatus should be inspected every 6 months in accordance with the following maintenance and inspection items.
* For repairs contact your local dealer or J. MORITA OFFICE.
1. Check that the battery does not seem to be losing its charge too quickly.
2. Check that pressing the Main Switch turns the unit on. After the unit is on, check that pressing the Main switch turns the
motor on and off. Check that the unit turns off when the Main Switch is pressed while the Select switch is being held down.
3. Check that pressing the Plus and Minus switches changes the memory number from M1 through M6.
4. Check that the settings for each memory can be changed.
5. Make sure the connection end of the motor handpiece is not damaged or dirty.
6. Make sure that the connection end of the contra angle is not damaged or dirty and that it can be securely connected to the
motor handpiece. Make sure that the file release button operates properly and that files can be securely installed.
When used with the Root ZX mini, make sure that the file electrode clips securely onto the file and that it is not damaged or
worn out.
7. When used with the Root ZX mini, touch the file with the contrary electrode and make sure that all the segments for the
meter light up properly.
Battery
*Inside motor handpiece
Transmission Cable
AR Oil
Power Supply Cord
Power Supply Cord
Contra Angle
File Electrode
(Code. No. 8450145)
(Code. No. 8450129)
(Code. No. 7503965)
(Code. No. 8450110)
(Code. No. 8450218)
(Code. No. 8450137)
(Code. No. 7503800)
(For Australia)
Handpiece Rest
(Code. No. 7505628 [9124055])
Operation 2016-08-22
Summary of Contents for Tri auto mini
Page 1: ...Cordless endodontic treatment motorized handpiece Tri Auto mini TR CM Operation Instructions...
Page 38: ......
Page 39: ......