background image

IPF

Page 14

Take the internal assemblies out of the housing by
grasping plastic removal tab and pulling straight out,
away from housing base.  Apply steady pull, disen-
gaging the pneumatic fittings from their ports in the
housing base.

Separate sub-assemblies by removing the single,
slotted-head screw in the side of section #1 of the
pneumatic block, which is shown in figure 7.

Cleaning the Mechanical
Subassembly

1. Place subassembly on work surface with zero

screw and span pot down.

2. Use syringe to fill flapper air passage with

trichloroethane (TCE).  See figure 7.

3. Clean nozzle air passage by gently passing

cleaning wire back and forth through opening in
subassembly base.

4. Soak several small strips of clean paper in TCE.

5. Set subassembly on its side, and slide one strip

between nozzle and flapper.  Carefully apply
slight pressure to flapper until it rests against
nozzle with soaked paper in between.

6. Maintain pressure while pulling paper out.

Repeat with other strips of paper until no residue
is transferred to paper.

7. Inspect air passage O-ring.  If damaged, contact

Moore Industries’ Customer Service Department
for replacement.

8. Use instrument air supply to dry and generally

“blow out” subassembly.  Put small amount of
alcohol in air passage, and set subassembly
aside.

Cleaning the Pneumatic Block

Refer to figure 7 to disassemble the IPF pneumatic
block, and when performing the following:

1. Place block on work surface with its two socket-

head screws facing up.  Use screwdriver to mark
one side of block with several small scratches for
use when re-assembling parts.

2. Dip cleaning wire in TCE, and use it to clean

orifice #1, the air passage between block and
mechanical subassembly (see figure 7).

Flush orifice with alcohol after cleaning.

3. Remove two socket-head screws that hold the

pneumatic block together.  Separate section #1
from block.

4. Use swabs dipped in TCE to clean internal

surfaces, and to remove any dirt particles or oil.
Flush all parts with alcohol after cleaning.

5. Push cleaning wire dipped in TCE through small

orifices on underside of section #1, in center
opening and along outside edge.  Use syringe
filled with alcohol to flush openings.

Set section #1 aside.

6. Remove spring, disk, and diaphragm from top of

section #2.  Inspect each for deterioration and
dirt.  Use compressed air to blow off all parts.

7. Remove diaphragm from bottom surface of

section if necessary.

8. Push cleaning wire dipped in TCE through brass

fitting in center of section.

9. Clean small orifices along edge of section (top

and bottom) with cleaning wire, and flush with
alcohol syringe.

Summary of Contents for IPF

Page 1: ...Field Mount Current to Pressure Transmitter May 2016 No 170 702 00 D 1992 by Moore Industries International Inc USER S MANUAL IPF...

Page 2: ...anada TOLLFREE 1 800 999 2900 UnitedKingdom FREEPHONE 0800525107 Australia TOLLFREE 008251928 16650 Schoenborn Street Sepulveda California 91343 U S A Tel 818 894 7111 Tlx 65 1322 FAX 818 891 2816 CON...

Page 3: ...eral third party industrial certifying agencies Contact your Moore Industries Sales Representative for the most current list of IPF certifications and approvals The unit has two electrical conduit dim...

Page 4: ...he combined effects of linearity hysteresis and repeatability For 3 27 psig output units or 10 50 mA input units error not to exceed 1 of span Linearity defined per SAMA standard PMC 20 1 1973 Ambient...

Page 5: ...using The example outlines the significance of each field of information in a typical IPF model number Calibration Every IPF is fully tested and calibrated at the factory prior to shipment Before inst...

Page 6: ...ore Industries but should be available in most labs or maintenance areas Calibration Setup To prepare the IPF for calibration disassemble the unit as shown in figure 1 and connect it to your calibrati...

Page 7: ...LOCK DETAIL I I N TYPE TERMINAL BLOCK DETAIL TERMINAL CONNECTIONS COVER ZERO ADJUST SCREW INTERNAL ASSEMBLY REMOVAL TAB SPAN POTENTIOMETER RUBBER RETAINING STRAP NE TYPE It is equipped with a slip clu...

Page 8: ...sing and to front panel of IPF For NE type units Route wiring through conduit opening in external sub housing terminal block assembly 5 Connect positive lead of adjustable current source to I terminal...

Page 9: ...to stabilize Calibration Procedure To perform the recommended bench check for the IPF first perform the setup as described in the preceding section The calibration procedure con sists of a basic chec...

Page 10: ...ed in an open loop this prevents compensating adjustments to the system input Mounting Figure 3 gives the IPF s outline dimensions The illustration also gives the dimensions of the available FR1 Optio...

Page 11: ...HREADS FOR WPM AND WPMP HOUSINGS FR1 OPTION APPROX 145 8 mm 5 7 in WHEN INSTALLED 9 14 mm 0 36 in DIAMETER 4 MOUNTING HOLES N TYPE TERMINAL CONNECTIONS 18 6 mm 0 72 in N TYPE 160 mm 6 28 in 94 mm 3 68...

Page 12: ...igure 4 Mounting the IPF on Vertical and Horizontal Pipe PNEUMATIC INSTRUMENT OR LOAD SUPPLY PORT OUTPUT PORT IPF CURRENT INPUT 4 20 OR 10 50 mA I I REGULATED INSTRUMENT AIR SUPPLY Figure 5 IPF Instal...

Page 13: ...to the 1 4 inch NPT female port on the side of the unit labeled SUPPLY Connect the output line to the 1 4 inch NPT female port labeled OUTPUT Refer to the unit front panel which identifies the IPF por...

Page 14: ...supply port and comes with a pressure gauge scaled in both psi 0 60 and bars 0 4 Operation Once its calibration has been checked or adjusted and the unit has been installed properly the IPF operates...

Page 15: ...must be cut to perform the maintenance cleaning Cut the wires as close to their butt connectors as possible Equipment Specifications Notes Instrument Quality Air or Nitrogen Supply Reduced to between...

Page 16: ...low out subassembly Put small amount of alcohol in air passage and set subassembly aside Cleaning the Pneumatic Block Refer to figure 7 to disassemble the IPF pneumatic block and when performing the f...

Page 17: ...13 Clean orifices and surfaces on section as before SECTION 1 HOLE ROLL IS UP ROLL IS DOWN HOLE RESTRICTOR HOLE SUPPLY SCREEN PNEUMATIC BLOCK FLAPPER NOZZLE MECHANICAL SUB ASSEMBLY Figure 7 IPF Intern...

Page 18: ...ned using an AWG 24 20 butt connector and heat shrink tubing available from Moore Industries Use of non specified connectors invalidates the group N certification Check to make sure that the spring is...

Page 19: ...the instructions found on the back cover of this manual If a problem is suspected with the IPF review the following steps 1 Verify that bench instruments used to take measurements have the proper rang...

Page 20: ...just ahead of each pressure regulator Water The amount of water in an air system depends on temperature pressure and the relative humidity of the air The amount of contamination therefore tends to var...

Page 21: ...IN OR MULTIPLE ON BRANCH LINES All Filters should have automatic drains and service indicators to show when elements need changing NOTE MAIN LINE TO OTHER BRANCHES PRESSURE REGULATOR PURIFIED AIR TO I...

Page 22: ...filtration to prevent oil contamination of the desiccant as well as to prevent desiccant fines from introducing new contamination ISA Specifications The Instrument Society of America standard ISA S73...

Page 23: ...R BREACH OF ANY WARRANTY CONCERNING THE GOODS OR SERVICES SHALL BE FOR THE COMPANY AT ITS OPTION TO REPAIR OR REPLACE THE GOODS OR SERVICES OR REFUND THE PURCHASE PRICE THE COMPANY SHALL IN NO EVENT B...

Reviews: