Portable RF communications equipment (radio devices) (including their acces-
sories such as antenna cables and external antennas) should not be used
closer than 30 cm (or 12 inches) from the parts and cables of the Docon 7
indicated by the manufacturer. Non-observance may result in a reduction of
the device's performance.
The requirements for the use in the areas of aviation, transportation and military have not
been considered as they have not been tested.
Electromagnetic
immunity test
/standard
IEC 60601-1-2
test level
Compliance level
Electromagnetic environment /
guidelines
Conducted RF
acc. to
IEC 61000-4-6
Docon 7 Standard
and Docon 7 RFID
Docon 7 Seal
3 V
rms
150 kHz to 30 MHz
3 V
rms
Portable and mobile RF communications
equipment should be used no closer to
any part of the Docon 7, including ca-
bles, than the recommended separation
distance calculated from the equation
applicable to the frequency of the trans-
mitter.
Recommended separation distance:
d = 1,2√P
for 80 MHz to 800 MHz
d = 2,3√P
for 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey
a)
should be less than
the compliance level in each frequency
range
b)
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
6 V
rms
in
ISM and
amateur radio fre-
quency bands be-
tween 150 kHz and
80 MHz
6 V
rms
Radiated RF acc.
to IEC 61000-4-3
Docon 7 Standard
Docon 7 RFID
Docon 7 Seal
10V/m
80 MHz to 2.7 GHz
Table
9
of
IEC 60601-1-2 Ed.4
Docon 7 Standard
Docon 7 RFID
Docon 7 Seal
10V/m
Table
9
of
IEC 60601-1-2 Ed.4