SpectraMax MiniMax Imaging Cytometer Installation Guide
26
5024111 C
Item
Description
1
Detection cartridge
2
Retaining clip unattached
3
Retaining clip attached
4
Retaining rod
5
Retaining rod
Figure 1-18: Detection Cartridge Retaining Clips and Retaining Rods
7. Install more detection cartridges, if desired.
8. Ensure that all the installed cartridges are evenly aligned and that the retaining
clips are securely fastened to the retaining rods on both sides of the detection
cartridge drawer.
9. Touch the detection cartridge drawer open/close button to close the detection
cartridge drawer.
10. Start the SoftMax Pro Software and connect to the instrument.
If the software is running and connected to the instrument, go to the
Operations
tab in the ribbon and click
Refresh
to let the software detect the installed detection
cartridges.
CAUTION!
To prevent dust from collecting inside the instrument, the
detection cartridge drawer should be open only for as long as necessary to
install or remove detection cartridges. Keep the detection cartridge drawer
closed whenever possible.