Installing and Setting Up the SpectraMax MiniMax Imaging Cytometer
5024111 C
15
To remove the transport locks:
1. Firmly pull on the front cover to remove it and then set it aside. The front cover is
held onto the front of the instrument by powerful magnets.
Tip:
It is easiest to remove the front cover by pulling from the bottom.
Figure 1-9: Detection Cartridge Drawer Cover Removed
2. Turn the knob on the detection cartridge drawer transport lock (2) counter-
clockwise until it is free of the threaded hole in the floor of the drawer
compartment.
Item
Description
1
Detection cartridge drawer
2
Detection cartridge drawer transport lock
Figure 1-10: Detection Cartridge Drawer Transport Lock
3. Gently slide the detection cartridge drawer (1) forward until it is outside of the
main instrument.
4. Lower the detection cartridge drawer transport lock (2) to remove it from the
detection cartridge drawer. Store the transport lock in the accessories tool box
included with the microplate reader.
5. Gently push the detection cartridge drawer back inside the instrument.