SpectraMax MiniMax Imaging Cytometer Installation Guide
14
5024111 C
Removing the Transport Locks
CAUTION!
The instrument can be damaged if the transport locks are not
removed before the instrument is powered on.
Transport locks are placed on the detection cartridge drawer and the microplate drawer
to protect the instrument from damage during shipping. The transport locks must be
removed before powering on the instrument.
This procedure requires the following tool:
Illustration
Part Number
Description
YW 000 006
Hex key, 2.0 mm
(included with the microplate reader)
Table 1-6: Required Tool
CAUTION!
Do not touch or loosen screws or parts other than those specifically
designated in the instructions. Doing so could cause misalignment and possibly
void the warranty.
CAUTION!
The front cover is held onto the front of the instrument by powerful
magnets. Keep magnetic storage devices or strips, such as hard drives, key
cards, and credit cards, away from the instrument covers.