![Molecular Devices SpectraMax MiniMax Installation Manual Download Page 11](http://html1.mh-extra.com/html/molecular-devices/spectramax-minimax/spectramax-minimax_installation-manual_1823758011.webp)
Installing and Setting Up the SpectraMax MiniMax Imaging Cytometer
5024111 C
11
6. Place one of the laser-interlock brackets (H) over the optics-access opening with the
interlock tab (I) on the bracket inserted into the corresponding alignment slot (J) on
the bottom of the microplate reader.
Item
Description
H
Laser-interlock brackets (2)
I
Interlock tabs (2)
J
Alignment slots (2)
K
Screws (4)
Figure 1-5: Laser-Interlock Brackets
7. Use the provided 2.5 mm hex key and two of the screws (K) provided with the
brackets to secure the laser-interlock bracket (H) to the bottom of the microplate
reader.
8. Attach the other laser-interlock bracket (H) with the interlock tab (I) inserted into
the corresponding alignment slot (J), and then use the other two provided screws
(K) to secure the bracket to the bottom of the microplate reader.
9. Lightly press on the interlock tabs (I) to make sure that they freely slide in and out
of the alignment slots (J).
Note:
The laser-interlock brackets are part of the hardware interlock that
prevents the laser module from turning on unless the cytometer is
installed. For the interlocks to work properly, the interlock tabs must freely
slide in and out of the alignment slots.