SpectraMax MiniMax Imaging Cytometer Installation Guide
8
5024111 D
Removing Shipping Protections
To prevent damage during shipping, four wing nuts secure the bolts in the corners, a
transport bracket with six blue retaining screws and three more blue retaining screws secure
the optics plate on the top of the cytometer, and a cover with three blue retaining screws
protects the optical components in the optics tower.
Optics Protection for Shipment
Description
A
4 Wing nuts and bolts
B
6 Transport bracket and retaining screws
C
3 Optics plate retaining screws
D
3 Optics cover and retaining screws
CAUTION!
Never touch the optic mirrors, lenses, filters, or cables. The optics are
extremely delicate, and critical to the function of the instrument.
Required Tool
Part Number
Description
5025413
3.0 mm hex key
To remove the shipping protections from the cytometer:
1. Use your hand to remove the four wing nuts (A) from the bolts attached to the corners
of the instrument. Do not remove the bolts.