SpectraMax Mini User Guide
42
5090339 A
Item
Description
Filter Based
Top Read
Dynamic Range
> 7 Decades
Sensitivity
2 pM (ATP)
Crosstalk
< 0.1%
Luminescence Performance
Item
Description
Reading chamber
Isothermal when temperature regulation is not enabled
Range
4°C above ambient to 45°C when temperature regulation enabled.
Note: To achieve temperature regulation at 45°C, the ambient
temperature must be > 20°C.
Resolution
Resolution ± 0.1°C
Accuracy
± 2.0°C for plate chamber
Temperature uniformity at
equilibrium
± 1°C at 37°C
Chamber warmup time
30-50 minutes (measured on air) after initiation of temperature
regulation
Temperature regulation
2 sensors
Drift
± 0.2°C (regulated)
Temperature regulation
diagnostics
Temperature regulation system is continuously monitored and
updated
Evaporation
Plate lid required to minimize evaporative cooling
Temperature Regulation