Chapter 2: Setting Up the Instrument
5090339 A
21
5. Grasp the filter cube by the stem and with the arrow pointing up, insert the filter cube into
the opening at the rear of the filter chamber. A magnet draws the filter cube into place and
holds the filter cube in the correct position within the instrument.
6. Close the filter chamber cover.
7. In the SoftMax Pro Software, on the Operations tab in the Ribbon, click
Info to display
the Instrument Information dialog to confirm successful filter cube installation. Click OK to
exit the Instrument Information dialog.