Software Configuration
SkanWasher 400 User Guide
21
When Sensor is enabled, the following parameters must be valid to start a
wash:
Air pressure must be close to the selected value.
A microplate must be on the microplate holder.
The waste reservoir must not be full.
Read Sensor
Instrument parameters can be verified using this Edit Option. Five sensor val-
ues are displayed for read only purposes. No changes can be made from this
screen. The five sensor values are from the following readings:
!! IMPORTANT:
When sensors
disabled, all three safety features
will be disengaged.
Air Pressure
Value is actual air pressure on internal pressure source
displayed in bar. Changes can be made through software
program. This sensor can be disabled through Edit
Options/Sensor On/Off.
Vacuum
Value is vacuum pressure, displayed in mbar. No
changes can be made through software program.
µP
Shows the status of the microplate lift 0-position sensor.
Displays in (+) or (-). This sensor can be disabled
through Edit Options/Sensor On/Off.
PL
Shows the status of the plate lift 0-position sensor. Dis-
plays in (+) or (-).
WF
Shows the status of the waste level sensor. Displays in
(+) or (-). This sensor can be disabled through Edit
Options/Sensor On/Off.