Software Configuration
SkanWasher 300 User Guide
18
Wash Step Selections
Each program can have up to 16 wash steps. There are seven wash step
selections available for each step. Press
[Enter]
to enter selection, and
use
¿À
arrow keys to change settings. Press
[Enter]
to select.
Wash Step
Description
Rinse Time
Set time in seconds from 0.1 to 10 seconds. Liquid will
be dispensed from “Rinse” port.
Disp Volume
Set dispense volume from 100-350 µl. Liquid will be
dispensed from “Inlet” port.
Soak/Pause
Set soak/pause time from 1 seconds to 2 minutes 50
seconds.
Wash volume
Set volume of wash from 100-900 µl. Liquid will be dis-
pensed from “Inlet” port.
Wash time
Set wash time from 0.1 to 10 seconds. Liquid will be dis-
pensed from “Inlet” port.
Aspirate
Set aspiration time from 2.0 - 10 seconds. Lift velocity is
varied with selected aspirate time. The lift velocity is
automatically set so that at the last second of the aspira-
tion, the aspirate probes are at the bottom of the well.
Aspirate times longer than 5 seconds will have the same
lift velocity as the 5-second setting.
Aspirate time = 2.0 seconds lift velocity approximately 1
second.
Aspirate time = 5 seconds lift velocity approximately 4
seconds.
End Wash
Set for last step of wash.
Table 1: Wash Step Selections
!! IMPORTANT:
To save changes,
Exit/Save must be selected after
program setting completed.
Summary of Contents for SkanWasher 300
Page 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Page 10: ......
Page 16: ......
Page 22: ......
Page 30: ......
Page 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Page 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Page 56: ......
Page 62: ......
Page 64: ......
Page 66: ......
Page 68: ......