Maintenance and Troubleshooting
5023399 A
23
To replace the Digidata 1550 digitizer fuses:
1.
Switch the power switch on the front of the instrument to the off
position.
2.
Unplug the power cord from the AC input connector.
3.
Press the carrier-release side-tabs towards each other and then
pull the fuse carrier to remove it from the instrument.
4.
Gently pull the old fuses from the carrier by hand.
5.
Gently place new fuses into the carrier by hand.
6.
Replace the fuse carrier. The carrier-release side-tabs click when
properly replaced.
7.
Plug the power cord into the AC input connector.
8.
Switch on power to the instrument.
Troubleshooting Problems and Solutions
This section provides specific troubleshooting advice on possible
problems and their solutions.
Before contacting Molecular Devices Technical Support regarding
problems, please review this section to see if you can address the
problem yourself. For more information, see
for faster resolution of your problem.
Isolating the Problem
The first rule in troubleshooting the digitizer or your software is to isolate
the problem.
1.
It is best to simplify your software configuration by turning off all
other programs. It might appear that there are no other
programs running in the background, but this might not be the
case. To see if other programs are running in the background, in
Windows, click on an empty place on the desktop and then press
Ctrl+Alt+Del
(holding these three keys down at the same time)
and then select
Task Manager
.
2.
On the
Applications
tab, close all unnecessary programs. Note
that virus checkers and automated Internet accesses can also
affect system performance.
Note:
If the instrument still does not power on after changing the fuses,
contact a Molecular Devices service engineer.