Appendix C: Operator
Appendix C: Operator
Performance Check
Performance Check
General information
General information
The CT100 and CT100HF Operator Performance Check is a series of procedures used to verify
the calibration of a CT100 or CT100HF unit. The following series of checks should be performed
after a unit has been newly calibrated to verify compliance to published specifications, but may
also be performed to determine if it requires recalibration.
The following series of steps are designed to allow a CT100 operator to completely verify the
properties of an individual CT100 unit to each published specification.
NOTE: If a CT100 fails any Operator Performance Check, it should be serviced.
Required equipment
Required equipment
Item
MOHR part number
50 Ω precision terminator
CT100-AC-ER50 (for CT100)
CT100-AC-ER50S (for CT100HF)
Capacitive calibrator
CT100-AC-C001
Connector, SMA male to BNC female
CT100-AC-ISMBF (for CT100HF)
Shorting cap
CT100-AC-IBS (for CT100)
CT100-AC-ISS (for CT100HF)
36-inch calibration cable
CT100-AC-W536 (for CT100)
CT100-AC-W536S (for CT100HF)
Male-male BNC connector
CT100-AC-IBMM (for CT100)
Additionally, a device capable of taking accurate pulse measurements with documented
calibration will be required.
Getting ready
Getting ready
Disconnect any cables from the front BNC connector. Connect the instrument to a suitable power
source, either a fully charged battery (internal or external) or the external 24V power supply
connected to a standard 3-prong AC source.