58
Appendix E. Communications
Figure E.1:
The
mogcmd
application, showing successful and unsuccessful
commands and queries.
E.2.1
Changing IP address
Depending on your network settings you may need to manually set
the
IP
address. This is most easily done via the front-panel interface
as detailed below. Once configured, these settings are stored in the
non-volatile memory of the unit and will be recalled in future.
1.
From the main menu, open Options
>
Ethernet Settings.
2.
Select
Static IP
and use the buttons to set the
IP
address of
the device as required.
3.
Select
Gateway
and set the gateway address as required.
4.
Select
DHCP
and set to
OFF
.
5.
Select
Restart ethernet
and press the
OK
button.
6.
The new
IP
address will be displayed in the display footer.
In some situations it may be necessary to power-cycle the device to
propagate ethernet changes.