1.
Connect to power
via the
USB
port or the
DC
barrel jack.
When
powering
with
USB
,
it
is
important
that
the
host
can
supply up to 600 mA. Some older computers may detect this as
a short-circuit and power down the device;
USB
-3.0 compliant
hubs are recommended.
2.
Power on
using the rocker switch on the rear.
3.
Connect fibre
using
the supplied
fibre patchcord.
Typically
this is
FC/APC
(green) at both ends, but can be
FC/PC
(black)
to
FC/APC
(green); in that case, the
FC/PC
black side must be
connected to the
FZW
.
Single-mode fibres are strongly preferred, although small-core
multi-mode fibres (up to 62
.
5
µ
m) can be used at the expense
of reduced accuracy.
4.
FSW
Fibre switch:
If using a fibre switch,
connect the out-
put fibre from the switch to the input fibre port on the
FZW
.
Electrically connect the
FSW
and
FZW
using the supplied M8
cable.
Connect your light source via fibre to one of the 4 or 8
input ports on the
FSW
.
5.
Input light
with the supplied free-space to fibre adapter. Typ-
ically the
FZW
only needs a few microwatts to operate, so high
coupling efficiency is not required.
The
saturation
is a measure of the power reaching the detector
The auto-exposure algorithm tunes the exposure time and sen-
sor gain to match the input power and optimise the measure-
ment rate.
Manual settings may be preferable, for example if
the laser power is rapidly fluctuating.
1
Absolute accuracy specifications are only valid when using single-mode fibres.
iii
Summary of Contents for FSW4
Page 1: ...Fizeau Wavemeter and Fibre Switcher FZW600 FSM2 FSW4 FSW8 Firmware v0 9 5 mogfzw v1 4 5 ...
Page 36: ...30 Chapter 5 PID locking ...
Page 50: ...44 Appendix A Specifications ...
Page 60: ...54 Appendix C Command language ...
Page 62: ...56 Appendix D Errors and troubleshooting ...
Page 67: ...F Dimensions and PCB 83 120 146 32 9 30V PID TRIG FSW Figure F 1 FZW chassis dimensions 61 ...
Page 68: ...62 Appendix F Dimensions and PCB 146 33 37 120 18 Figure F 2 FSW fibre switch dimensions ...
Page 71: ......