M I L L E N N I U M M E D I C A L P R O D U C T S , I N C .
34
For an ANCHOR SHACKLE and SCREW PIN
•
Install the SCREW PIN through the SHACKLE and the LIFT POINT hole on the BOOM.
•
Screw the PIN fully into the threaded part of the SHACKLE to secure the SCALE to the BOOM
LIFT POINT. Make sure the threads of the PIN are fully engaged in the SHACKLE. Assemble
the LIFTING BAR or STRETCHER SLING SYSTEM to the bottom of the SCALE. Use the same
shoulder bolt/self locking nut assembly that secured the LIFTING BAR to the BOOM LIFT
POINT. Tighten the self-locking nut fully.
•
The LIFTING BAR or STRETCHER SLING SYSTEM should be free to rotate about its
connection to the bottom of the SCALE.
For a PUSH TO RELEASE PIN
•
Insert the PUSH TO RELEASE PIN fully through the bar on the top of the SCALE and the LIFT
POINT on the BOOM.
•
Assemble the LIFTING BAR or STRETCHER SLING SYSTEM
to the bottom of the SCALE.
Use the same PUSH TO RELEASE PIN that secured the LIFTING BAR to the BOOM LIFT
POINT.
•
The LIFTING BAR should be free to rotate about its connection to the bottom of the SCALE.
3. OPERATE THE SCALE AND WEIGH THE PATIENT
Three steps are necessary to weigh a patient:
a)
Viewing the SCALE's digital readout panel, select the appropriate readout in
kg
or
lbs
by pressing
the Left Button under the front panel of the scale. The units of
kg
or
lbs
will be displayed on the left
side of the digital SCALE panel.
b)
Assemble the SLING to the LIFTING BAR or STRETCHER SLING SYSTEM (including FABRIC
STRETCHER SLING) and zero out the weight of these parts. Depress the "Zero" button (right
button) momentarily. "0" will pan across the digital display for a few seconds then the display will
read "0.0".
c)
Remove the SLING or FABRIC STRETCHER SLING and then position the patient into the sling on
the bed and attach the SLING to the LIFTING BAR or the FABRIC STRETCHER SLING to the
STRETCHER SLING ARCH. Carefully lift the patient free of any support such as the bed or wheel
chair. Once the motion is stabilized press the "Operate" button (center button) to weigh the patient.
The digital readout will show an increasing number and the weight will be displayed in a short
period of time. The "Operate" button can be depressed several times while stabilized to repeat the
weight measurement. The weight will be displayed for about 60 seconds.
4.
IMPORTANT
POINTS
a)
MAKE SURE ALL ASSEMBLY HARDWARE IS TIGHT AND SECURE.
b)
Make sure the SCALE readout faces away from the MAST when installed.