background image

 

MKS Instruments UK Ltd 

V2000-C/E Hardware Manual – SP102005.102 March 2013

 

 

 

28 

 

Heating and Baking 

The RGA chamber runs at a temperature of 70°C and the inlet valve UniBloc assembly at 90°C.  Preset, limited 
power is applied to these heater jackets any time the turbopump is on.  This improves gas response time by 
reducing the residence time of gases on the walls, especially polar species.  From the time of starting the 
turbopump, allow at least 2 hours for temperature equilibration.  During this time, background gas levels will vary, 
significantly. 
 
Before you can start to use the system to its full potential, you will need to run it for sufficient time to allow the 
background in the RGA vacuum chamber to drop.  
Baking the system will significantly reduce this time. This should be done after the system has run for at least an 
hour to allow the pressure in the system to drop below 1x10

-5

 mBar. We recommend baking the RGA vacuum 

chamber for at least 20 hours.  
 
Optimum bake-out is achieved by having the inlet valve open with the process chamber at the normal operating 
pressure of very clean and dry inert gas, or by having all inlet valves closed (the V3 purge valve will be open). 
The quadrupole should be running with the filament on, but MUST be using the faraday detector. 
 

DO NOT

 use the multiplier detector during baking! 

DO NOT

 use the electron multiplier detector within the first two hours after terminating baking! 

 
 
 
 

CAUTION

 

The Electron Multiplier (SEM) MUST NOT be operated at 

temperatures above 90

o

C.  Do NOT bake with the SEM selected.  

After baking ends, wait at least 2 hours before using the SEM! 

 

 
 

The electron multiplier will be seriously damaged if it is operated at temperatures above 90

0

C! 

 

No damage is caused to the multiplier by high temperatures provided it is not switched on. 

 

The only remedy when a multiplier has been damaged due to being operated at higher temperatures is to 

replace it. 

 
 
Baking can be started (and stopped) directly at the RVC2 with the BAKE button.  It can also be controlled 
through the RVC status and control panel of Process Eye Professional. 
 
The total pressure should gradually start to rise and you should bake the system at least until that pressure starts 
to drop. 
 
It is useful to create timed bakeout recipes to run with the Faraday detector and turn the bake Off after a preset 
time so the system can cool down unattended. The default Bake period is 8 hours. 
 
To improve the background further it is recommended that you run and degas both filaments. The amount of 
time spent in reducing the background peaks depends entirely on the application and is left to the discretion of 
the customer. If you switch the system off it will vent to atmosphere introducing water vapour and you will have to 
bake again.  Venting to clean dry nitrogen or argon is strongly recommended.  Note that the RGA chamber will 
vent, partially, to the supplied purge gas whenever the turbopump is off. 
 
 
 

Warning, exposed metal surfaces reach high temperatures during bake-out. Allow an adequate 

cooling down period before handling

 

 

Summary of Contents for Vision 2000-C

Page 1: ...Vision 2000 C E Manual SP102005 102 March 2013...

Page 2: ...ticularly relating to any omissions or inaccuracies you may discover You can send your comments to MKS Instruments Spectra Products Cowley Way Weston Road Crewe Cheshire CW1 6AG United Kingdom 44 0 12...

Page 3: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 3...

Page 4: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 4...

Page 5: ...MC and safety issues The vacuum system on which the analyser RF head is mounted must be earthed to a protective earth preferably to the same protective earth as the control unit Operation The equipmen...

Page 6: ......

Page 7: ...ecting the foreline and SurgeProtector valve 19 Heater jackets 20 MicroVision2 installation 20 ELECTRICAL INSTALLATION 21 Equipment rack 22 Interconnecting cables 22 CONNECTING TO THE PROCESS TOOL 23...

Page 8: ...these boxes is described below WARNING WARNING boxes are used where failure to observe the instructions could result in personal injury or death CAUTION CAUTION boxes are used where failure to observ...

Page 9: ...ar panel of the Microvision2 refers to Read all instructions carefully before use The control and signal ports are designed for connection to MKS Instruments accessories via MKS Instruments cables The...

Page 10: ...mm H R C ceramic 250V AC characteristic T and compliant with IEC 127 Details of fuse types and ratings can be found in the RVC2 manual and printed on the rear panel Electrical Connections The Vision 2...

Page 11: ...om the tool foreline when the RGA system pump is off and when the tool foreline pressure is above 3 torr WARNING The RGA system must never be operated without exhausting to a hazardous gas scrubbing s...

Page 12: ...uments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 12 System Options Fig 1 3 Valve Etch Configuration Fig 2 4 Valve Fast Response LPCVD Configuration Fig 3 4 Valve High Pressure CVD Confi...

Page 13: ...r local MKS Instruments facility or sales service representative Carefully unpack the various parts of your Vision 2000 C E system Again check for any signs of damage Find the shipment report and chec...

Page 14: ...n 2000C E has makes little difference to its installation On the face of the UniBloc inlet valve that connects to the RGA chamber is a ceramic CIS closed ion source coupling Exchanging an inlet valve...

Page 15: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 15 Dimensions 490mm 225mm 265mm 340mm 166mm...

Page 16: ...depending on working pressure 722 Baratron Gauge 2 LoVacOK 890B Gauge 3 UniBloc inlet assembly 4 Purge regulator assembly 3 5 to 8Bar Argon or dry Nitrogen supply required 5 System Interface Module 6...

Page 17: ...e Manual SP102005 102 March 2013 17 Item Ref Description 8 Pneumatic Manifold assembly 60 80 psig dry air 9 Fore line connecting flange 10 Cable loom supporting bracket 8 9 10 Pneumatics supply Use ad...

Page 18: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 18 Item Ref Description 15 V1 16 V2 17 V3 18 V4 19 UniBloc mounting face 20 16 15 17 18 19...

Page 19: ...g response time can be improved by using a length of 1 8 OD PTFE tubing from the point to be sampled to the face of the UniBloc Be sure that the tubing material is compatible with the process gases to...

Page 20: ...must be strictly followed Heater jackets The Vision 2000C E is supplied with two heater jackets fitted one on the main RGA chamber and another around the inlet valve assembly There is no need to remov...

Page 21: ...s and regulations The electrical installation should be carried out after the vacuum system installation Please follow the next sections in sequence Electrical specification See the RVC2 and Microvisi...

Page 22: ...n protective conduit e g plastic trunking that provides a good degree of protection according to the location of installation All cables are labelled at both ends All the connectors have been chosen t...

Page 23: ...100 120 Volts AC 50 60 Hz 15 Amps 220 240 Volts AC 50 60 Hz 13 Amps LINE BROWN NEUTRAL BLUE EARTH GREEN YELLOW Computer connection Vision 2000C E systems are used in conjunction with MKS Instruments...

Page 24: ......

Page 25: ...mputer software The operation of the vacuum system controller is fully described in the Remote Vacuum Controller manual The pumps are switched on from the Remote Vacuum Controller window or from the R...

Page 26: ...r will indicate when the pump is up to speed and the Turbopump indicator in the RVC window will change from yellow to green Note that the turbopump controller will be programmed to a set point of 95 o...

Page 27: ...unt recipe in Process Eye and check that the probe gas is set to mass 4 for helium or the appropriate mass if you are using another gas As the Vision 2000C E uses a Microvision2 there can be an audio...

Page 28: ...CAUTION The Electron Multiplier SEM MUST NOT be operated at temperatures above 90 o C Do NOT bake with the SEM selected After baking ends wait at least 2 hours before using the SEM The electron multip...

Page 29: ...ween the two sequential orifice stages O2 1 and O2 2 The inlet pressures that can be sampled properly are determined by this configuration and the size of the orifices Again differential pumping is ob...

Page 30: ......

Page 31: ...the two flying leads coming from the gauge connector and measure the voltage Ensure that the process chamber pressure is below 1milli torr then check the meter reading If the indicated voltage is gre...

Page 32: ...oval For system with compression springs note the following tightening procedure tighten these screws until the top of the screw head is flush with the top surface of the ceramic coupling springs will...

Page 33: ...tron multiplier SEM detector will also require replacement In extreme cases this can be as soon as 3 months but 1 2 years is more typical For non corrosive gas sampling SEM performance is often accept...

Page 34: ...f 6 o clock the standard analyser alignment pip orientation is at 9 o clock Remove the six x 28 bolts Carefully withdraw the analyser from the vacuum chamber Leave the old copper gasket in place until...

Page 35: ...rned delays in providing this completed form will lead to delays in the servicing of the instrument Securely package all items to be returned using the original packaging where possible and send to th...

Page 36: ...ease ensure that we have this information before we receive the equipment A copy should also be given to the carrier FAILURE TO COMPLETE THIS FORM OR COMPLY WITH THE PROCEDURE WILL LEAD TO DELAYS IN S...

Page 37: ...ese substances Action to be taken in the event of human contact or spillage I hereby confirm that the only toxic or hazardous substances that the equipment specified above has been in contact with are...

Reviews: