PAS Trolley Manual
LP101017 Rev 1.02
Page 28
Poor vacuum due to primary pump performance - refer to MKS
Turbo pump overheating. Check turbo cooling fan.
Problem with turbo controller/turbo pump - refer to MKS.
7. Cold Cathode gauge does not switch on.
Turbo pump not at speed.
Check fuse FS4.
Problem with gauge controller - refer to MKS.
8. Cold Cathode gauge does not indicate pressure.
Check connection to Cold Cathode gauge head.
Gauge head faulty/contaminated - refer to MKS.
9. RGA does not work - green LED on Electronics unit not lit.
Check RGA power supply cable connection.
Vacuum not low enough to trip low pressure trip point. Check
system for leaks.
Fault with RGA power supply - refer to MKS.
10. RGA does not work - green LED on Electronics lit.
Problem with electronics unit - refer to MKS.
11. No power to Laptop computer.
Check Laptop mains cable.
Check fuse FS5.
12. Heater Indicator does not illuminate.
Check fuse FS5.
13. Bakeout Heater indicator illuminates, heater controller does not indicate.
Problem with heater controller - refer to MKS
Summary of Contents for Spectra 138A
Page 10: ...PAS Trolley Manual LP101017 Rev 1 02 Page ii This page is deliberately left blank ...
Page 17: ...PAS Trolley Manual LP101017 Rev 1 02 Page 9 This page is deliberately left blank ...
Page 19: ...PAS Trolley Manual LP101017 Rev 1 02 Page 11 ...
Page 20: ...PAS Trolley Manual LP101017 Rev 1 02 Page 12 This page is deliberately left blank ...
Page 21: ...PAS Trolley Manual LP101017 Rev 1 02 Page 13 ...
Page 22: ......
Page 25: ...PAS Trolley Manual LP101017 Rev 1 02 Page 17 ...
Page 26: ......
Page 34: ......
Page 38: ......
Page 41: ...PAS Trolley Manual LP101017 Rev 1 02 Page 33 ...
Page 42: ......
Page 44: ...PAS Trolley Manual LP101017 Rev 1 02 Page 36 This page is deliberately left blank ...