Signalmode
(extra window for channel 1 and channel 2)
SIGMODE:
For controllers the parameter
SIGMODE
determines the source of the setpoint which shall be
applied the the respective channel. For meters this parameter determines to which value the
setpoint output is set.
INDEP
(Independent) Setpoint is set either by the front panel settings or
by remote host computer. Manual setpoint adjust can be done
either in the setup submenu or in a window in the operational level
(if the window is configured for displaying setpoint).
OFF
Setpoint output voltage is set to a fixed voltage of about – 0,5 V .
This ensures that the integrator of a controller connected to this
channel does not wind up thus opening the control valve
unintendendly. The digital output connector CHANNEL1 or
CHANNEL2, pin 4 is held to LOW. This activates the input
Valve
Close
of MFC’s, forcing the control valve to be closed.
METER
Setpoint output voltage is set to zero volts. Valve override line via
pin 4 of the respective channel connector is set to HIGH.
RTD
Range-Turndown (RTD) allows operation of a Baratron type 120.
The digital output pin 4 of connector CHANNEL1 or CHANNEL2 is
set to LOW when the signal of the type 120 is falls below 1 V. This
enables the PR4000F to display also the lower pressure decades
in full reolution.
Refer to section
CONTROL
detailed instructions.
SLAVE
The setpoint of a slave channel is derived from the other channel’s
signal. This allows the control of gas mixing systems with a fixed
flow ratio.
EXTRN
(external setpoint)
The channel’s setpoint is derived by an external analog voltage
source, connected to connector
EXTERN
.
SIGMODE1 INDEP
SCL1 10.000 mbar
Summary of Contents for PR 4000 F
Page 8: ...List of Figures and Tables PR4000F viii This page left blank...
Page 12: ...Safety information PR4000F 4 This page left blank...
Page 50: ...Chapter 4 Operation PR4000F 42 This page left blank...
Page 64: ...Appendix A Specifications PR4000F 56 This page left blank...
Page 66: ...Appendix B Error Messages PR4000F 58 This page left blank...
Page 68: ...Appendix C Default Configuration PR4000F 60 This page left blank...