Cables for combination with the PR4000F
For pressure transducers or in-line pressure controllers
type or series
MKS-Cable Type
120
121
622, 623, 624, 625, 223, 122A
621, 626, 627, 628, 640, 641, 127, 128, 722A (with 15-pin type D
connector)
722 (9-pin type D connector)
722 (terminal block)
For mass flow meters (MFM) or mass flow controllers (MFC)
with 15-pin type D connector:
179, 1179, 2179, 1479, 1259, 2259, 258, 358, 1359, 558, 1559,
M100, 1640
with 9-pin type D connectorr:
1179, 2179, 1479, M200, M330
CBE 120-96-3M
CBE 112-14-3M
CBE 112-2-3M
CBE 259-5-3M
CBE 700-1-3M
CBE 700-99-3M
CBE 259-5-3M
CBE147-12-3M
Table 2: Standard Interconnecting Cables
Note
Flow controllers with 9-pin connector do not have the „Valve Close“ input
(remotely closing of the control valve).
Generic Shielded Cable Description
MKS offers a full line of cables for all MKS equipment. Should you choose to manufacture your own
cables, follow the guidelines listed below:
1. The cable must have a
braided
shield, covering all wires. Neither aluminum foil nor spiral
shielding will be as effective; using either may nullify regulatory compliance.
2. The connectors must have a metal case which has direct contact to the cable’s shield on the
whole circumference of the cable. The inductance of a flying lead or wire from the shield to the
connector will seriously degrade the shield’s effectiveness. The shield should be grounded to
the connector before its internal wires exit.
3. With very few exceptions, the connector(s) must make good contact to the device’s case
(ground). “Good contact” is about 0.01 ohms; and the ground should surround all wires.
Contact to ground at just one point may not suffice.
Summary of Contents for PR 4000 F
Page 8: ...List of Figures and Tables PR4000F viii This page left blank...
Page 12: ...Safety information PR4000F 4 This page left blank...
Page 50: ...Chapter 4 Operation PR4000F 42 This page left blank...
Page 64: ...Appendix A Specifications PR4000F 56 This page left blank...
Page 66: ...Appendix B Error Messages PR4000F 58 This page left blank...
Page 68: ...Appendix C Default Configuration PR4000F 60 This page left blank...