background image

 

The Ariel User Manual 

Rev 5, 01/04/21 

 

10 

 

Chapter 5: Pulsed Power 

5.1: Physical Principle 

The core of the Ariel is a high power copper thermopile disk that can measure short exposures of high 
power lasers. 

An internal, un-calibrated silicon photodiode detects the backscattered light to measure the exposure time. 

Average power is then calculated by, 

𝑃 =

𝐸

∆𝑡

 ,

 where 

P

 is power, 

E

 is energy, and 

Δt

 is the time interval. 

5.2: Limitations 

The photodiode's sensitivity depends on wavelength, meaning that the minimum power that may be 
detected by it also depends on the wavelength. The photodiode is not sensitive beyond 1100nm. Please 
check the specifications for details. 

2.94µm and 10.6µm lasers are not supported by the photodiode.  

5.3: Display 

The calculated power will be displayed in large digits. The energy will be 
displayed in small digits on the left side, and the pulse width will be displayed 
in small digits on the right side of the screen. 

On

 

the pulsed power screen, a blinking 'RDY' message (meaning “READY”) 

will appear on the top right corner to indicate that the Ariel is ready to 
measure a pulse. 

5.4: Errors 

Situation 

Device displays 

Log note 

No trigger detected 

No trigger! 

Pulse is too long 

OVER 

Pulse too long! 

Pulse is too short 

UNDER 

Pulse too short! 

Table 3: Pulsewidth errors

 

In a situation where the pulse width is not detected but the energy is measured, the device will display the 
energy but not the calculated peak power. The peak power can be calculated manually if the pulse width 
is known, using the equation above. 

Figure 9: Pulse screen

 

Summary of Contents for Ophir Ariel

Page 1: ...POB 45021 Jerusalem Israel 9145001 972 2 548 7458 Fax 972 2 548 7481 E mail ophir mksinst com www ophiropt com Rev5 01 04 21 Instruction Manual Ariel User Manual...

Page 2: ...be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying and recording or by any information storage or retrieval system except as may be expressly per...

Page 3: ...Features 8 Chapter 4 Wavelength options 9 Chapter 5 Pulsed Power 10 5 1 Physical Principle 10 5 2 Limitations 10 5 3 Display 10 5 4 Errors 10 Chapter 6 Using the Ariel 11 6 1 Power on 11 6 2 Power of...

Page 4: ...r 9 Figure 7 Empty top 9 Figure 8 Diff screen 9 Figure 9 Pulse screen 10 Figure 10 The Ariel front view 11 Figure 11 Measurement screen controls 11 Figure 12 Status line 11 Figure 13 Pulse measurement...

Page 5: ...ce and repair to ensure that all safety features are maintained SERVICE BY QUALIFIED PERSONNEL ONLY Operating personnel must not remove instrument covers Component replacement and internal adjustments...

Page 6: ...seconds The Ariel then measures the energy and duration of the laser pulse and calculates the power The Ariel is designed to work without water cooling The Ariel is versatile and durable with a dust...

Page 7: ...20W 800 1100nm 10W 1100nm not available c Maximum Beam Incidence Angle Without diffuser 30 degrees for 12mm Gaussian beam With diffuser 25 degrees for 10mm Gaussian beam d Reproducibility 1 Power Ran...

Page 8: ...angular pulse is assumed for this calculation b May be used at 550 900nm with decreased accuracy and higher reflection up to 10 c Use without window or diffuser The sensor does not measure pulse width...

Page 9: ...The Ariel User Manual Rev 5 01 04 21 7 Figure 2 The Ariel drawing...

Page 10: ...ed before switching on the laser USB C cable used for charging and as a data cable for StarLab Wall charger Case has a separate sleeve for the window diffuser while not in use USB rubber plug used in...

Page 11: ...user 2 94 m 2 94 m 32mm The top cover should be left open Figure 7 Empty top 10 6 m 10 6 m 32mm The top cover should be left open Table 2 Wavelengths options Make sure to screw the window diffuser tig...

Page 12: ...lasers are not supported by the photodiode 5 3 Display The calculated power will be displayed in large digits The energy will be displayed in small digits on the left side and the pulse width will be...

Page 13: ...socket that cannot provide minimum 500mA will prevent the Ariel from charging properly and may cause damage to the charger 6 4 Display and navigation The display has a status line on the top This lin...

Page 14: ...rformance to extend duration before cool down is needed and to shorten cool down duration Warning Overheating the Ariel or using power density higher than stated in the specifications may cause damage...

Page 15: ...via Bluetooth Disable Enable disables or enables the Bluetooth connection Clear All clears the Bluetooth cache Use this option only if you encounter pairing difficulties Temperature units Toggles the...

Page 16: ...PC See more details in the StarLab User Manual 7 2 StarViewer The StarViewer app may be used on an Android device in order to communicate with the Ariel using Bluetooth wireless communication Note See...

Reviews: