Chapter Five: Theory of Operation
How To Read the Flow Output Directly
33
How To Read the Flow Output Directly
If your readout equipment does not provide a GCF (gas correction factor) potentiometer, you
must add a proper input divider to read the flow signal directly.
Using the values from the example above, and assuming a 1 Volt F. S. readout, the flow signal
must be divided 2.44 to 1.00 Volts.
2.44 V
10 K
1.00 V (= 100 sccm)
14.4 K
Use 1% RN60C
resistors, or better
Figure 8: Wiring Diagram for an Input Divider
Note
MKS Types 246, 247, and 260 provide gas correction factor adjustment.