Chapter 2
22
Mini-Convectron Module with RS-485
Instruction Manual - 275545
ConFlat Flange
To minimize the possibility of leaks with ConFlat-type flanges, use high
strength stainless steel bolts and a new, clean OFHC copper gasket. Avoid
scratching the seal surfaces. To avoid contamination, do not use nonmetal
gaskets.
Finger tighten all bolts. Use a wrench to continue tightening
1
/
8
turn at a
time in crisscross order, e.g., 1, 4, 2, 5, 3, 6, 4 until all flanges are in contact.
After contact, further tighten each bolt about
1
/
16
turn.
2.9
Dimensions
• Dimensions are in cm (in.).
• All dimensions are nominal. For tolerances, contact the Granville-Phillips
Customer Service Department.
Figure 2-3
Dimensions