Appendix A: Product Specifications
Type 128 Performance Specifications
25
Type 128 Performance Specifications
Accuracy
0.1 Torr
±
1 to 25K Torr
±
0.5% of Reading
±
temperature coefficient
0.25% of Reading
±
temperature coefficient
CE Compliance
Electromagnetic Compatibility
2
EMC Directive 89/336/EEC
Resolution
0.01% of Full Scale
Response Time
<16 milliseconds
Temperature Coefficient
Zero
0.1 Torr
1 Torr to 25K Torr
Span
0.015% F.S./
°
C
0.005% F.S./
°
C (over 20 to 70
°
C)
0.02% Reading/
°
C (over 20 to 70
°
C)
2
An overall metal braided shielded cable, properly grounded at both ends, is required during use.
Summary of Contents for Baratron 127A
Page 3: ...110660 P1 Rev E 6 97 MKS Baratron Type 127A and Type 128A Absolute Pressure Transducers ...
Page 8: ...List of Figures and Tables vi ...
Page 16: ...Customer Support Chapter One General Information 8 This page intentionally left blank ...
Page 30: ...General Information Chapter Five Maintenance 22 This page intentionally left blank ...