
Chapter Four: Operation
How To Zero the Type 127/128 Transducer
19
Chapter Four: Operation
How To Zero the Type 127/128 Transducer
You must zero the transducer initially when you install it into your system and periodically
thereafter. The zero must be set by adjusting the zero potentiometer.
Note
Make sure that the unit is fully stabilized before attempting to adjust the
zero.
1. Pump down the unit, with power on, to a pressure less than the resolution (0.01% F.S.)
for approximately two to four hours (eight hours are required for the 0.1 Torr range).
Table 3, page 19, lists the maximum pressure for each full scale range.
2. Adjust the zero by turning the Zero potentiometer, located on the top of the unit.
Refer to Figure 5, page 15, for the location of the Zero pot.
The zero adjustment may also be made at the front panel of an MKS supplied readout/power
supply unit, if used.
Maximum Pressure Reading for Zero Adjustment
Full Scale Range
Maximum Pressure for Zero Adjustment
0.1 Torr
5 x 10
-6
Torr
1 Torr
5 x 10
-5
Torr
2 Torr
1 x 10
-4
Torr
10 Torr
5 x 10
-4
Torr
100 Torr
5 x 10
-3
Torr
1000 Torr
5 x 10
-2
Torr
5000 Torr
2.5 x 10
-1
Torr
10,000 Torr
5 x 10
-1
Torr
15,000 Torr
7.5 x 10
-1
Torr
20,000 Torr
1 Torr
25,000 Torr
1.25 Torr
Table 3: Maximum Pressure Reading for Zero Adjustment
Summary of Contents for Baratron 127A
Page 3: ...110660 P1 Rev E 6 97 MKS Baratron Type 127A and Type 128A Absolute Pressure Transducers ...
Page 8: ...List of Figures and Tables vi ...
Page 16: ...Customer Support Chapter One General Information 8 This page intentionally left blank ...
Page 30: ...General Information Chapter Five Maintenance 22 This page intentionally left blank ...
Page 38: ...Index 30 V Vibration 13 W Wetted material 5 24 26 Z Zero 13 19 21 ...