How To Adjust the Analog Set Point Value
Chapter Six: Remote Operation
76
How To Adjust the Analog Set Point Value
The 651 controller is capable of accepting one analog set point through the I/O connector on the
rear panel. The analog set point value is expressed as a % of full scale, using the following
formula:
analog set point = (analog set point voltage / full scale voltage)
For example, if the 651 unit is configured for 5 Volt full scale input, and the actual input applied
is 3 Volts, the analog set point value will be 60%. Similarly, if the 651 unit is configured for 10
Volt full scale input, and the actual input applied is 5 Volts, the analog set point value will be
50%.
The system responds to an analog set point voltage by immediately storing the analog set point
value. The system pressure does not change however, unless the analog set point is the active set
point. Refer to How To Determine the Active Set Point, page 72, for more information.
How To Set the Analog Set Point Full Scale Range
The analog set point can be configured for 5 Volt or 10 Volt full scale input.
1. To change the analog set point full scale range, issue the command:
A value
ENTER
where value is 0 for 5 Volt range; 1 for 10 Volt range.
2. To check the analog set point full scale range, issue the request:
R24
ENTER
The 651 controller will send the following response:
A
value
where
value
is 0 for 5 Volt range; 1 for 10 Volt range.
Summary of Contents for 651C
Page 12: ...List of Figures x...
Page 14: ...List of Tables xii...
Page 34: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Page 110: ...Digital Logic Control Chapter Six Remote Operation 96 This page intentionally left blank...
Page 170: ...Safety Information Page 6 of 8 This page intentionally left blank...