Appendix F: Command and Request Reference
Command Reference
129
RS-232 Command Summary (Continued)
Command
Function
D3
Select set point C
D4
Select set point D
D5
Select set point E
D6
Select analog set point
E value
Sensor range value
0 = 0.1
10 = 1000
1 = 0.2
11 = 5000
2 = 0.5
12 = 10000
3 = 1
13 = 1.33
4 = 2
14 = 2.66
5 = 5
15 = 13.33
6 = 10
16 = 133.3
7 = 50
17 = 1333
8 = 100
18 = 6666
9 = 500
19 = 13332
F value
Pressure units value
0 = Torr
1 = mTorr
2 = mbar
3 = µbar
4 = kPa
5 = Pa
6 = cmH
2
O
7 = inH
2
O
Note: The F command merely assigns a label to the
pressure units. It does not convert pressure readings.
Pressure readings are % of F. S.
G value
Sensor voltage range value
0 = 1 Volt
1 = 5 Volts
2 = 10 Volts
O
Open valve
C
Close valve
H
Hold valve
Table 26: RS-232 Command Summary
(Continued on next page)