Chapter Six: Remote Operation
How To Configure the Sensor Parameters
67
How To Change the Sensor Range
The 651 controller is initially configured to work with a 100 Torr pressure sensor. If your sensor
covers a different range, issue the command:
E value
ENTER
where value is a valid sensor range. Refer to Table 15 for the complete list of valid sensor
ranges.
Sensor Range Values
Torr
millibar
0
=
0.1
13
=
1.33
1
=
0.2
14
=
2.66
2
=
0.5
15
=
13.33
3
=
1
16
=
133.3
4
=
2
17
=
1333
5
=
5
18
=
6666
6
=
10
19
=
13332
7
=
50
8
=
100*
9
=
500
10
=
1000
11
=
5000
12
=
10000
* initial value
Table 15: Sensor Range Values
Note
Pressure readings are reported as % of full scale (F.S.), where full scale is
the sensor range value shown in the table. For example, if the actual
pressure is 10 Torr for a 10 Torr F.S. unit, then the 651 controller reports
a pressure reading value = 100 (for 100%).
If the pressure is 10 Torr for a 100 Torr F.S. unit, then the 651 controller
reports a pressure reading value = 10 (for 10%).