Safety precaution
CAUTION
●
Devices and accessories used together with this product
•
Before using other devices or accessories, thoroughly read the instruction manual
of the devices and make sure that the operating table is not affected adversely.
Before fitting on accessories from third party companies, contact your distributor
or Mizuho. Some accessories cannot be fitted on.
•
While operating the operating table, check the position of other devices or the
accessories used with them. They may come in contact with each other during the
operation, operating table, devices, and/or accessories may get damaged.
●
Cleaning and disinfection
•
After using the operating table, make sure to follow the steps below to clean up
and disinfect the operating table.
1. Turn off the power and disconnect the power cord from the medical grade outlet.
2. Detach all the mattresses from the operating table.
3. Use a lint-free cloth soaked with proper volume of disinfectant to wipe off the
upper, side, and back side of the mattresses.
4. As with step 3, disinfect the surfaces of the tables and side rails.
5. Wipe off the operating table with a clean dry cloth 15 minutes after disinfecting it.
•
Make sure to use Mizuho authorized disinfectants. The disinfectants are as shown
below.
1
Sodium hypochlorite 0.1% (halogen containing compound)
2
Sodium thiosulfate
3
Chlorhexidine (chlorhexidine gluconate 0.5%)
4
Benzalkonium chloride (invert soap 10%)
5
Povidone iodine
6
Ethanol 80%
7
Hydrogen peroxide
8
Saline (0.9%)
9
Isopropyl alcohol (IPA)